Xantrex Technology XFR 2800 manual Local and Remote Sensing, Sense Wiring

Models: XFR 2800

1 90
Download 90 pages 12.58 Kb
Page 55
Image 55
Local and Remote Sensing

Installation

Local and Remote Sensing

Local and Remote Sensing

Use connections at the rear panel J10 sense connector to configure the power supply for local or remote sensing of output voltage. See Figure 2.11 for a drawing of the sense connector.

Sense Wiring

WARNING

There is a potential shock hazard at the sense connector when using a power supply with a rated output greater than 40 V. Select wiring with a minimum insulation rating equivalent to the maximum output voltage of the power supply for use as local sense jumpers or for remote sense wires. For example, select TEW-105, 105 °C, 600 V wiring for use with a model 600 V, 4 A power supply. Ensure that connections at the load end are shielded to prevent contact with hazardous voltages.

For lowest noise performance, use shielded pair wiring of 16 to 24 AWG for remote sense lines. Strip wires 0.26 in. (6.5 mm) and insert securely as with any wire clamp connector.

Local We ship the power supply with the rear panel J10 sense connector jumpered for local Sensing sensing of the output voltage. See Table 2.5 for the list of connector functions and a

description of local sense connections. With local sensing, the output voltage is regulated at the output terminals (or bus bars). This method does not compensate for voltage losses in the load lines, so it is recommended only for low current applications or applications for which precise load regulation is not essential.

Note:When using local sense connections, use the largest practical load wire size to minimize the effects of line impedance on the regulation of the supply.

Release 2.1

53

Page 55
Image 55
Xantrex Technology XFR 2800 manual Local and Remote Sensing, Sense Wiring