Tool Required: Phillips #2 screwdriver

Calibration

Calibration Setup

Removing Option Card If there is a GPIB or RS-232 option card (PCB) installed inside the power supply, you will have to remove it to access the calibration potentiometers.

Tool Required: Phillips #2 screwdriver

1.Turn OFF the power supply. Disconnect AC Power. Remove the cover.

2.Remove 4 screws from their standoffs inside the power supply to release the option card.

3.Remove 2 screws at the rear panel to release the option card’s sub-plate.

4.Unplug the ribbon cable connectors from the option card. Lift out card.

Figure 5.1

Programming and Monitoring Calibration Locations

(Top view.)

Release 2.1

81

Page 83
Image 83