Calibration
Calibrating for Readback Accuracy
Calibrating for Readback Accuracy
Output
Voltage
Monitor
Circuit
Calibration
The factory calibrates the offset and range of the output voltage and current monitor circuits to within 1% for the default
See Section for instructions to remove the cover and any option card. Follow all procedure steps in the sequence given.
1.Ensure that the power supply is turned OFF. Disconnect any load.
2.Set SW1 switch 5 OPEN to select
3.Connect a DVM across the power supply output to read the output voltage. Connect a second DVM between the rear panel J2 connector terminals 10 (output voltage monitor) and 9 (voltage monitor signal return).
4.Turn the power supply ON.
5.Set the power supply output voltage to 1% of the
6.Adjust the output voltage monitor offset (R397) until the monitor DVM reads 1% of the readback scale. (50 mVdc for
7.Set the power supply output voltage to the
8.Adjust the output voltage monitor range (R387) until the monitor DVM reads 100% of the readback scale. (5 Vdc for
Repeat this procedure until the monitor signal is adjusted to power supply specifications (see Section 1) or to the accuracy that your application requires.
84 | Operating Manual for XFR 2.8kW Series Power Supply |