Xantrex Technology XFR 2800 manual Calibrating for Programming Accuracy

Models: XFR 2800

1 90
Download 90 pages 12.58 Kb
Page 84
Image 84

Calibration

Calibrating for Programming Accuracy

Calibrating for Programming Accuracy

Voltage

Programming

Circuit

Calibration

The factory calibrates the offset and range of the voltage and current programming circuits to within 1% for the default 0-10 Vdc programming signals. You may need to recalibrate when you use 0-5 Vdc programming or when you switch back to 0-10 Vdc programming after previously calibrating for 0-5 Vdc programming.

See “Calibration Setup” on page 79 for instructions to remove the cover and any option card. Follow all procedure steps in the sequence given.

1.Ensure that the power supply is turned OFF. Disconnect any load.

2.Connect the program source between J2 connector terminals 6 (output voltage program input) and 5 voltage program signal return).

3.Connect the DVM across the power supply output.

4.Turn the power supply ON.

5.Apply 1% of program voltage.

6.Adjust the voltage program offset potentiometer (R472) until the DVM reads 1% of the model-rated output voltage.

7.Apply 100% of program voltage.

8.Adjust the voltage program range potentiometer (R462) until the DVM reads the model-rated output voltage.

Repeat this procedure until output levels are adjusted to power supply specifications (see Section 1) or to the accuracy that your application requires.

82

Operating Manual for XFR 2.8kW Series Power Supply

Page 84
Image 84
Xantrex Technology XFR 2800 manual Calibrating for Programming Accuracy, Voltage Programming Circuit Calibration