Current Limit

Programming

Circuit

Calibration

Calibration

Calibrating for Programming Accuracy

1.Ensure that the power supply is turned OFF. Disconnect any load.

2.Connect the program source between J2 connector terminals 8 (output current limit programming input) and 7 (current program signal return).

3.Connect the shunt and DVM across the power supply output.

4.Turn the power supply ON.

5.Apply 1% of program voltage.

6.Adjust the current limit program offset (R469) until the shunt/DVM indicates 1% of the model-rated output current. See Note.

7.Apply 100% of program voltage.

8.Adjust the current limit program range (R459) until the shunt/DVM indicates the model-rated output current. See Note.

Repeat this procedure until output levels are adjusted to power supply specifications (see Section 1) or to the accuracy that your application requires.

Note:Calculate the required DVM reading using V= IR where V is the DVM reading, I is the current, and R is the DC shunt resistance.