Table 1-2. Supplemental Characteristics (continued)

Fuse: The ac input is protected by a fuse located in a module on the rear panel; 0.5AM for l00/120 Vac input; 0.25AM for 220/240 Vac input.

Maximum VA: 60

Peak Inrush Current: 2.5 A (typical)

PROGRAMMABLE SLEW RATE: (For any given input transition, the time required will be either the total slew time or a minimum transition time, whichever is larger. The minimum transition time increases when operating with input currents under 1 AM (6060B) or 0.2 AM (6063B) and decreases with input currents over 20 A (6060B) or 2 A (6063B). The following are typical values; ± 25% tolerance.)

Current Slew Rate:

Model 6060B (Ac performance specified from 3 to 60 V)

Rate #

High Range Step

Low Range Step

Transition Time

1

1 A/ms

0.1 A/ms

8.0 ms

2

2.5 A/ms

0.25 A/ms

3.2 ms

3

5 A/ms

0.5 A/ms

1.6 ms

4

10 A/ms

1 A/ms

800 μs

5

25 A/ms

2.5 A/ms

320 μs

6

50 A/ms

5 A/ms

160 μs

7

0.1

A/μs

10 A/ms

80 μs

8

0.25

A/μs

25 A/ms

32 μs

9

0.5

A/μs

50 A/ms

16 μs

10

1

A/μs

0.1 A/μs

12 μs

11

2.5

A/μs

0.25 A/μs

12 μs

12

5

A/μs

0.5 A/μs

12 μs

Model 6063B (Ac performance specified from 3 to 240 V)

Rate #

High Range Step

Low Range Step

Transition Time

1

0.17

A/ms

17

A/s

8.0 ms

2

0.42

A/ms

42

A/s

3.2 ms

3

0.83

A/ms

83

A/s

1.6 ms

4

1.7 A/ms

0.17

A/ms

800 μs

5

4.2 A/ms

0.42

A/ms

320 μs

6

8.3 A/ms

0.83

A/ms

160 μs

7

17

A/ms

1.7 A/ms

80 μs

8

42

A/ms

4.2 A/ms

32 μs

9

83

A/ms

8.3 A/ms

20 μs

10

0.17

A/μs

17 A/ms

20 μs

11

0.42

A/μs

42 A/ms

16 μs

12

0.83

A/μs

83

A/ms

16 μs

 

 

 

 

 

 

General Information 15

Image 15
Contents Operating Manual Agilent Part No Microfiche Part No Updated AprilCertification Safety Summary Manufacturer’s Declaration Safety SummarySymbol Description HerstellerbescheinigungPrinting History Table of Contents Local Operation Remote OperationCalibration Considerations for Operating in Constant Resistance Mode General Information What’s In This ManualReader Path OptionsSpecifications Safety Requirements6060B 6063B Regulation Constant Resistance Mode RangesModes Continuous Mode Pulsed ModeCurrent Readback Supplemental Characteristics 6060B 6063B ResolutionTemperature Coefficient Bandwidth Maximum Readback CapabilityRate # High Range Step Low Range Step Transition Time Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Current Slew RateResistance Slew Rate OvershootModel 6063B Range Transient Current Level Width 6060B 6063B 10%With unit on With unit offPage Operation Overview IntroductionFront Panel Description Remote ProgrammingModes of Operation Local/Remote ControlProgrammable Features Constant Current CC ModeSoftware Current Limit Triggered Current LevelConstant Resistance CR Mode Slew RateImmediate Resistance Level Constant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Transient Operation Immediate Voltage LevelTriggered Voltage Level Transient Voltage LevelPulsed Transient Operation Hpsl Command DescriptionSelects the external trigger input source Selects the external trigger inputSelects pulsed transient operation Sets pulse width to 1 millisecondSlew Rate And Minimum Transition Time Triggering a preset levelTriggering a transient pulse TogglingRisetime Transition Limitation Input Current, Voltage, and Power MeasurementTransition Times and Slew Rates Short On/OffStatus Reporting Saving and Recalling SettingsReading Remote Programming Errors Input On/OffOvercurrent Protection FeaturesResetting Latched Protection OvervoltageReverse Voltage Control ConnectorOverpower OvertemperatureFault Monitor OutputsExternal Programming Input Port On/Off Inspection Location and Cooling Turn-On CheckoutCheck Line Voltage Line Voltage Fuse Agilent Part No Voltage Select SwitchesTurn-On/Selftest Connect The Power CordGpib Errors Input Errors Display Description ErrorPower Test Gpib ConnectorController Connection AWG Rear Panel Connectors and SwitchesWire Size Strip back Gpib AddressInput Binding Post Control Connector Common Pins 3+S and -S IM and VMSense Switch Trigger ConnectorApplication Connections Maximum Wire Lengths to Limit Voltage Drops Parallel ConnectionsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Local Operation Front Panel Controls and Indicators DescriptionControls and Indicators Description Function Keys Local Control Overview Using The Function Keys ∙ Short onTurning the Input On/Off Recommended Programming Sequence Key Display Setting the Mode of OperationSetting CC Values CC Programming Ranges FunctionExamples Setting CR Values Examples Setting CV Values Function Key Display Range of Values Continuous Pulse Train Programming RangesShorting The Input Using The System KeysDisplaying Error Codes Setting The Electronic Load’s Gpib AddressChanging Wake-up Settings Factory Default Settings Function 6060B 6063BRecalling the Factory Default Values Enter Enter/Output StatementsGpib Address OutputOutput 705 MEASCURR? Sending a Remote CommandRemote Programming Commands Getting Data BackSlew Rates ModesRanges Transient levelsRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CC Mode Example CV Mode ExampleCR Mode Example Pulsed Transient Operation Example Continuous Transient Operation ExampleConstant Current CC Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Remote Programming Ranges FunctionTransient Operation Constant Resistance CR6060B Constant Voltage CVPage Equipment Required for Calibration Equipment Characteristics Recommended ModelExample Programs Equipment RequiredCalibration Commands Example Program Calibration FlowchartsCalibration Information 6060B 6063B Supply Shunt Values SettingsCalibration Flowchart Calibration Flowchart Calibration Flowchart END if Program ListingPause WaitPrint Voltage Calibration Pause SubendElse Line ExplanationPage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates