Agilent Technologies 5951-2826 Selects the external trigger input, Selects toggled operation

Page 26

HPSL Command

Description

TRIG:SOUR EXT

Selects the external trigger input.

TRAN:MODE PULS

Selects pulsed transient operation.

CURR 5

Sets main current level to 5 amps.

CURR:TLEV 10

Sets transient current level to 10 amps.

TRAN:TWID .001

Sets pulse width to 1 millisecond.

TRAN ON

Turns on transient operation.

Figure 2-5 shows the waveform that would result in this pulsed transient operation example. The Electronic Load starts conduction at the main current level setting (5 amps). When the transient mode is turned on and an external trigger signal is received, the input level starts increasing at a rate determined by the slew rate. When the value specified by the transient level setting (10 amps) is reached, it stays there for the remainder of the time determined by the pulse width setting

(1 millisecond). After this time has elapsed, the input level decreases to the main level again at the rate specified by the slew setting and remains there until another trigger is received. Any triggers that occur during the time the transient level is in effect will be ignored.

Figure 2-5. Pulsed Transient Operation

Toggled Transient Operation

Toggled transient operation causes the load input to alternate between two predefined levels as in continuous operation except that the transient points are controlled by explicit triggers instead of the internal transient generator. As in pulsed transient operation, the trigger signal can be an external trigger signal, the GPIB GET function, the *TRG command, or the TRIG command. Note that toggled transient operation can only be programmed via the GPIB (TRAN:TOGG command); it cannot be programmed at the front panel.

In this example, assume that the CC mode is active, the slew rate is at the factory default setting (maximum rate), an external trigger input signal is connected to the Electronic Load’s rear panel, and the applicable transient operation parameters have been set as follows:

HPSL Command

Description

TRIG:SOUR EXT

Selects the external trigger input source.

TRAN:MODE TOGG

Selects toggled operation.

CURR 5

Sets main current level to 5 amps.

CURR:TLEV 10

Sets transient current level to 10 amps.

TRAN ON

Turns on transient operation.

26 Operation Overview

Image 26
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Herstellerbescheinigung Safety SummarySymbol Description Manufacturer’s DeclarationPrinting History Table of Contents Calibration Local OperationRemote Operation Considerations for Operating in Constant Resistance Mode Options What’s In This ManualReader Path General InformationSafety Requirements SpecificationsPulsed Mode Regulation Constant Resistance Mode RangesModes Continuous Mode 6060B 6063BCurrent Readback Temperature Coefficient Supplemental Characteristics 6060B 6063BResolution Maximum Readback Capability BandwidthCurrent Slew Rate Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Rate # High Range Step Low Range Step Transition TimeModel 6063B Range Transient Current Level Resistance Slew RateOvershoot With unit off 6060B 6063B 10%With unit on WidthPage Remote Programming IntroductionFront Panel Description Operation OverviewConstant Current CC Mode Local/Remote ControlProgrammable Features Modes of OperationTriggered Current Level Software Current LimitImmediate Resistance Level Constant Resistance CR ModeSlew Rate Transient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Transient Voltage Level Immediate Voltage LevelTriggered Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSets pulse width to 1 millisecond Selects the external trigger inputSelects pulsed transient operation Selects the external trigger input sourceToggling Triggering a preset levelTriggering a transient pulse Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesInput On/Off Saving and Recalling SettingsReading Remote Programming Errors Status ReportingOvervoltage Protection FeaturesResetting Latched Protection OvercurrentOvertemperature Control ConnectorOverpower Reverse VoltageExternal Programming Input FaultMonitor Outputs Port On/Off Inspection Check Line Voltage Location and CoolingTurn-On Checkout Voltage Select Switches Line Voltage Fuse Agilent Part NoError Connect The Power CordGpib Errors Input Errors Display Description Turn-On/SelftestController Connection Power TestGpib Connector Gpib Address Rear Panel Connectors and SwitchesWire Size Strip back AWGInput Binding Post Control Connector IM and VM Pins 3+S and -S CommonTrigger Connector Sense SwitchApplication Connections Zero-Volt Loading Connections Maximum Wire Lengths to Limit Voltage DropsParallel Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Turning the Input On/Off Using The Function Keys∙ Short on Recommended Programming Sequence CC Programming Ranges Function Setting the Mode of OperationSetting CC Values Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Output Enter/Output StatementsGpib Address EnterGetting Data Back Sending a Remote CommandRemote Programming Commands Output 705 MEASCURR?Transient levels ModesRanges Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CR Mode Example CC Mode ExampleCV Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleRemote Programming Ranges Function Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Constant Current CCConstant Voltage CV Constant Resistance CR6060B Transient OperationPage Equipment Required Equipment Characteristics Recommended ModelExample Programs Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Wait Program ListingPause END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates