Agilent Technologies 5951-2827, 5951-2826 manual Modes, Ranges, Transient levels, Slew Rates

Page 67

1.Modes

The CC, CR, and CV values can be programmed whether or not the associated mode is active. If the input is turned on, all of the applicable values will take effect at the input when the associated mode is selected.

2.Ranges

Changing the CC or CR programming range can cause the present settings to be automatically adjusted to fit within the new range. See Setting CC Values and Setting CR Values in Chapter 4. During a range change, the input will go through a non-conducting state to minimize overshoots.

3.Transient levels

The transient CC or CV level must be set to a higher level than the respective main level. In the low range, the transient CR level must be set to a higher level than the main CR level. In the middle and high ranges, the transient CR level must be set to a lower level than the main CR level.

4.Slew Rates

The CC slew rate is programmed in amps/second. There are 12-steps for each of the two current ranges (low and high). The Electronic Load automatically selects one of the 12 steps that is closest to the programmed value. The CV slew rate is programmed in volts/second. There are 12-steps within the voltage range. The Electronic Load automatically selects one of the 12 steps that is closest to the programmed value. In the low range, the CR slew rate is programmed in volts/second instead of ohms/second. Whatever value is programmed for the CV slew rate is also used for CR. In the middle and high ranges, the CR slew rate is programmed in amps/second. Whatever value is programmed for the CC slew rate is also used for CR.

5.Programmable Current Protection (CURR:PROT)

The programmable current limit is in effect for any mode of operation (not just the CC mode). When programmable current protection is enabled, and the programmed current limit and time delay are exceeded, the module’s input will be turned off.

6.Measurement Overload (OVLD)

If the input voltage exceeds the maximum measurement capability of the load, an overload (OVLD) condition will be indicated in the return values that resulted from a MEAS:VOLT? or MEAS:POW? query sent to the associated channel. The MEAS:POW? query will return an overload indication if either voltage or current has exceeded the module’s maximum measurement capability since power is calculated from voltage and current. Overload is indicated by the value 9.9E + 37 instead of the normal voltage or power readings. This is the IEEE 488.2 value for positive infinity.

Remote Operation 67

Image 67
Contents Operating Manual Agilent Part No Microfiche Part No Updated AprilCertification Safety Summary Manufacturer’s Declaration Safety SummarySymbol Description HerstellerbescheinigungPrinting History Table of Contents Remote Operation Local OperationCalibration Considerations for Operating in Constant Resistance Mode General Information What’s In This ManualReader Path OptionsSpecifications Safety Requirements6060B 6063B Regulation Constant Resistance Mode RangesModes Continuous Mode Pulsed ModeCurrent Readback Resolution Supplemental Characteristics 6060B 6063BTemperature Coefficient Bandwidth Maximum Readback CapabilityRate # High Range Step Low Range Step Transition Time Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Current Slew RateOvershoot Resistance Slew RateModel 6063B Range Transient Current Level Width 6060B 6063B 10%With unit on With unit offPage Operation Overview IntroductionFront Panel Description Remote ProgrammingModes of Operation Local/Remote ControlProgrammable Features Constant Current CC ModeSoftware Current Limit Triggered Current LevelSlew Rate Constant Resistance CR ModeImmediate Resistance Level Triggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Transient Operation Immediate Voltage LevelTriggered Voltage Level Transient Voltage LevelPulsed Transient Operation Hpsl Command DescriptionSelects the external trigger input source Selects the external trigger inputSelects pulsed transient operation Sets pulse width to 1 millisecondSlew Rate And Minimum Transition Time Triggering a preset levelTriggering a transient pulse TogglingRisetime Transition Limitation Input Current, Voltage, and Power MeasurementTransition Times and Slew Rates Short On/OffStatus Reporting Saving and Recalling SettingsReading Remote Programming Errors Input On/OffOvercurrent Protection FeaturesResetting Latched Protection OvervoltageReverse Voltage Control ConnectorOverpower OvertemperatureMonitor Outputs FaultExternal Programming Input Port On/Off Inspection Turn-On Checkout Location and CoolingCheck Line Voltage Line Voltage Fuse Agilent Part No Voltage Select SwitchesTurn-On/Selftest Connect The Power CordGpib Errors Input Errors Display Description ErrorGpib Connector Power TestController Connection AWG Rear Panel Connectors and SwitchesWire Size Strip back Gpib AddressInput Binding Post Control Connector Common Pins 3+S and -S IM and VMSense Switch Trigger ConnectorApplication Connections Parallel Connections Maximum Wire Lengths to Limit Voltage DropsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Local Operation Front Panel Controls and Indicators DescriptionControls and Indicators Description Function Keys Local Control Overview ∙ Short on Using The Function KeysTurning the Input On/Off Recommended Programming Sequence Key Display Setting the Mode of OperationSetting CC Values CC Programming Ranges FunctionExamples Setting CR Values Examples Setting CV Values Function Key Display Range of Values Continuous Pulse Train Programming RangesShorting The Input Using The System KeysDisplaying Error Codes Setting The Electronic Load’s Gpib AddressChanging Wake-up Settings Factory Default Settings Function 6060B 6063BRecalling the Factory Default Values Enter Enter/Output StatementsGpib Address OutputOutput 705 MEASCURR? Sending a Remote CommandRemote Programming Commands Getting Data BackSlew Rates ModesRanges Transient levelsRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CV Mode Example CC Mode ExampleCR Mode Example Pulsed Transient Operation Example Continuous Transient Operation ExampleConstant Current CC Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Remote Programming Ranges FunctionTransient Operation Constant Resistance CR6060B Constant Voltage CVPage Equipment Required for Calibration Equipment Characteristics Recommended ModelExample Programs Equipment RequiredCalibration Commands Example Program Calibration FlowchartsCalibration Information 6060B 6063B Supply Shunt Values SettingsCalibration Flowchart Calibration Flowchart Calibration Flowchart END if Program ListingPause WaitPrint Voltage Calibration Pause SubendElse Line ExplanationPage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates