Agilent Technologies 5951-2826 Setting The Electronic Load’s Gpib Address, Displaying Error Codes

Page 62

Setting The Electronic Load’s GPIB Address

Before you can program the Electronic Load remotely via a GPIB computer, you must know its GPIB address. You can

find this out by pressing . The Electronic Load’s GPIB address will be displayed; for example "ADDRESS 5". The Electronic Load is shipped from the factory with its address set to 5.

If you want to leave the address set at 5, you can return to the metering mode by pressing the Meter key.

If you want to change the address, you can enter a new value. Any integer from 0 to 30 can be selected. For example, to change the address to 12 press:

.

This new address will remain set and will not be lost when power is cycled. Note that the Address setting is not affected by the Save and Recall functions described below.

Displaying Error Codes

Remote programming errors are indicated when the Err annunciator is on. To display the error code(s), first return to local

control by pressing .

To display an error code, press (blue shift key) (shifted).

Errors are recorded in a list and are displayed in the order in which they occurred. Each time the shifted Error key is pressed, an error code is displayed. Once an error is displayed, it is removed from the error list. "ERROR 0" indicates there are no errors present and will be displayed when all errors in the list have been displayed. The error codes are negative numbers in the range from - 100 to - 499. Refer to the Agilent Electronic Loads Programming Reference Guide for a description of the error codes.

Saving and Recalling Settings

The Electronic Load’s settings (mode, input state, current levels, resistance levels, etc.) can be saved and then recalled for use in various test setups. The complete list of parameters that can be saved and recalled are the same parameters as listed in Table 4-6.

The present settings of all parameters can be saved in a specified storage register (0 to 6) using the Save (shifted Recall) key. At a later time, you can recall the settings from the specified register using the Recall key.

For example, you can store the present settings in register 2 by pressing (blue shift key) (shifted) .

You can change the Electronic Load’s settings as required and then return to the settings stored in register 2 by pressing .

Settings stored in registers 1 through 6 will be lost when the Electronic Load’s power is cycled. When power is turned off and then on again, each of these registers (1 through 6) will be set to the "wake-up" values. The "wake-up" values are stored in register 0 and can be set to any values you desire (see Changing Wake-up Settings).

The main advantage in using internal registers 1 through 6 is that it simplifies the repetitive programming of different settings. The Save key can be used in conjunction with the Input on/off key to store settings while the input is off. The Recall key can be used at a later time to recall desired settings while the input is turned on.

62 Local Operation

Image 62
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Herstellerbescheinigung Safety SummarySymbol Description Manufacturer’s DeclarationPrinting History Table of Contents Calibration Local OperationRemote Operation Considerations for Operating in Constant Resistance Mode Options What’s In This ManualReader Path General InformationSafety Requirements SpecificationsPulsed Mode Regulation Constant Resistance Mode RangesModes Continuous Mode 6060B 6063BCurrent Readback Temperature Coefficient Supplemental Characteristics 6060B 6063BResolution Maximum Readback Capability BandwidthCurrent Slew Rate Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Rate # High Range Step Low Range Step Transition TimeModel 6063B Range Transient Current Level Resistance Slew RateOvershoot With unit off 6060B 6063B 10%With unit on WidthPage Remote Programming IntroductionFront Panel Description Operation OverviewConstant Current CC Mode Local/Remote ControlProgrammable Features Modes of OperationTriggered Current Level Software Current LimitImmediate Resistance Level Constant Resistance CR ModeSlew Rate Transient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Transient Voltage Level Immediate Voltage LevelTriggered Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSets pulse width to 1 millisecond Selects the external trigger inputSelects pulsed transient operation Selects the external trigger input sourceToggling Triggering a preset levelTriggering a transient pulse Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesInput On/Off Saving and Recalling SettingsReading Remote Programming Errors Status ReportingOvervoltage Protection FeaturesResetting Latched Protection OvercurrentOvertemperature Control ConnectorOverpower Reverse VoltageExternal Programming Input FaultMonitor Outputs Port On/Off Inspection Check Line Voltage Location and CoolingTurn-On Checkout Voltage Select Switches Line Voltage Fuse Agilent Part NoError Connect The Power CordGpib Errors Input Errors Display Description Turn-On/SelftestController Connection Power TestGpib Connector Gpib Address Rear Panel Connectors and SwitchesWire Size Strip back AWGInput Binding Post Control Connector IM and VM Pins 3+S and -S CommonTrigger Connector Sense SwitchApplication Connections Zero-Volt Loading Connections Maximum Wire Lengths to Limit Voltage DropsParallel Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Turning the Input On/Off Using The Function Keys∙ Short on Recommended Programming Sequence CC Programming Ranges Function Setting the Mode of OperationSetting CC Values Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Output Enter/Output StatementsGpib Address EnterGetting Data Back Sending a Remote CommandRemote Programming Commands Output 705 MEASCURR?Transient levels ModesRanges Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CR Mode Example CC Mode ExampleCV Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleRemote Programming Ranges Function Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Constant Current CCConstant Voltage CV Constant Resistance CR6060B Transient OperationPage Equipment Required Equipment Characteristics Recommended ModelExample Programs Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Wait Program ListingPause END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates