Agilent Technologies 5951-2827, 5951-2826 manual Using The System Keys, Shorting The Input

Page 61

a.Set the main CC level to 0.5 amps, the transient CC level to 1 amp, and the slew rate to 0. 0025 A/μs. See examples under Setting CC Values.

b.Turn on CC mode by pressing:

2.Set frequency to 5 kHz by pressing:

3.Set duty cycle to 25% by pressing:

(blue shift key) (shifted)

4.Turn on transient operation by pressing:

5.Note that the Tran annunciator is on.

Shorting The Input

The Electronic Load can simulate a short circuit across its input. The short circuit can be toggled on/off by pressing

.

When the input is shorted the message "SHORT ON" win be displayed. The short on/off change uses the slew rate setting of the active mode and range. Turning the short off returns the input to the previously programmed values and returns the display to the metering mode. Note that "INPUT OFF" takes precedence over "SHORT ON".

Pressing the Short on/off key with certain user applications may cause damage to the equipment being tested, which may result in personal injury. Contact your Agilent Sales and Service office if you need to have the Short on/off key disabled.

Resetting Latched Protection

The Electronic Load includes overvoltage "OV", overpower "OP", and overtemperature "OT" protection features as well as a software overcurrent limit protection feature (remotely programmable only) that latch when they are tripped. The protection shutdown "PS" and voltage fault "VF" conditions also latch when tripped. The Prot annunciator on the front

panel goes on when any of the above features are tripped. To reset any of these protection features, press .

Note

The condition that caused the protection feature to trip must be removed or it will trip again as soon as it

 

is reset. Also, if OT occurs, the Electronic Load must have sufficiently cooled down in order for the

 

to take effect.

 

 

Using The System Keys

These keys consist of Local, Address, Error (shifted Address key), Recall, Save (shifted Recall key), and the blue shift key. The Local key and the Shift key have already been discussed. The remaining SYSTEM keys are described in the following paragraphs.

Local Operation 61

Image 61
Contents Operating Manual Agilent Part No Microfiche Part No Updated AprilCertification Safety Summary Symbol Description Safety SummaryHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Remote Operation Local OperationCalibration Considerations for Operating in Constant Resistance Mode Reader Path What’s In This ManualOptions General InformationSpecifications Safety RequirementsModes Continuous Mode Regulation Constant Resistance Mode RangesPulsed Mode 6060B 6063BCurrent Readback Resolution Supplemental Characteristics 6060B 6063BTemperature Coefficient Bandwidth Maximum Readback CapabilityMaximum VA Peak Inrush Current 2.5 a typical Supplemental CharacteristicsCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeOvershoot Resistance Slew RateModel 6063B Range Transient Current Level With unit on 6060B 6063B 10%With unit off WidthPage Front Panel Description IntroductionRemote Programming Operation OverviewProgrammable Features Local/Remote ControlConstant Current CC Mode Modes of OperationSoftware Current Limit Triggered Current LevelSlew Rate Constant Resistance CR ModeImmediate Resistance Level Triggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Triggered Voltage Level Immediate Voltage LevelTransient Voltage Level Transient OperationPulsed Transient Operation Hpsl Command DescriptionSelects pulsed transient operation Selects the external trigger inputSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a transient pulse Triggering a preset levelToggling Slew Rate And Minimum Transition TimeRisetime Transition Limitation Input Current, Voltage, and Power MeasurementTransition Times and Slew Rates Short On/OffReading Remote Programming Errors Saving and Recalling SettingsInput On/Off Status ReportingResetting Latched Protection Protection FeaturesOvervoltage OvercurrentOverpower Control ConnectorOvertemperature Reverse VoltageMonitor Outputs FaultExternal Programming Input Port On/Off Inspection Turn-On Checkout Location and CoolingCheck Line Voltage Line Voltage Fuse Agilent Part No Voltage Select SwitchesGpib Errors Input Errors Display Description Connect The Power CordError Turn-On/SelftestGpib Connector Power TestController Connection Wire Size Strip back Rear Panel Connectors and SwitchesGpib Address AWGInput Binding Post Control Connector +S and -S Pins 3IM and VM CommonSense Switch Trigger ConnectorApplication Connections Parallel Connections Maximum Wire Lengths to Limit Voltage DropsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Local Operation Front Panel Controls and Indicators DescriptionControls and Indicators Description Function Keys Local Control Overview ∙ Short on Using The Function KeysTurning the Input On/Off Recommended Programming Sequence Setting CC Values Setting the Mode of OperationCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Function Key Display Range of Values Continuous Pulse Train Programming RangesShorting The Input Using The System KeysDisplaying Error Codes Setting The Electronic Load’s Gpib AddressChanging Wake-up Settings Factory Default Settings Function 6060B 6063BRecalling the Factory Default Values Gpib Address Enter/Output StatementsOutput EnterRemote Programming Commands Sending a Remote CommandGetting Data Back Output 705 MEASCURR?Ranges ModesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CV Mode Example CC Mode ExampleCR Mode Example Pulsed Transient Operation Example Continuous Transient Operation ExampleOutput 705TRAN on Output 705INPUT on Hpsl Command Short FormRemote Programming Ranges Function Constant Current CC6060B Constant Resistance CRConstant Voltage CV Transient OperationPage Example Programs Equipment Characteristics Recommended ModelEquipment Required Equipment Required for CalibrationCalibration Commands Example Program Calibration FlowchartsCalibration Information 6060B 6063B Supply Shunt Values SettingsCalibration Flowchart Calibration Flowchart Calibration Flowchart Pause Program ListingWait END ifPrint Voltage Calibration Pause SubendElse Line ExplanationPage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates