Agilent Technologies 5951-2826, 5951-2827 manual Index

Page 92

 

INDEX (continued)

software current limit

33

status reporting

32, 55

system keys

........................................ 52, 63

 

T

toggled transient operation

29

transient current level

23

transient operation

26, 62

transient resistance level

25

transient voltage level

26

transition time

29

trigger connector

44

triggered current level

23

triggered operation

29

triggered resistance level

24

triggered voltage level

26

 

V

voltage fault

55

 

W

wake-up settings

65

wire lengths

47

 

Z

zero-volt loading

46

92 Index

Image 92
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Safety Summary Symbol DescriptionHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Calibration Local OperationRemote Operation Considerations for Operating in Constant Resistance Mode What’s In This Manual Reader PathOptions General InformationSafety Requirements SpecificationsRegulation Constant Resistance Mode Ranges Modes Continuous ModePulsed Mode 6060B 6063BCurrent Readback Temperature Coefficient Supplemental Characteristics 6060B 6063BResolution Maximum Readback Capability BandwidthSupplemental Characteristics Maximum VA Peak Inrush Current 2.5 a typicalCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeModel 6063B Range Transient Current Level Resistance Slew RateOvershoot 6060B 6063B 10% With unit onWith unit off WidthPage Introduction Front Panel DescriptionRemote Programming Operation OverviewLocal/Remote Control Programmable FeaturesConstant Current CC Mode Modes of OperationTriggered Current Level Software Current LimitImmediate Resistance Level Constant Resistance CR ModeSlew Rate Transient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Immediate Voltage Level Triggered Voltage LevelTransient Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSelects the external trigger input Selects pulsed transient operationSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a preset level Triggering a transient pulseToggling Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesSaving and Recalling Settings Reading Remote Programming ErrorsInput On/Off Status ReportingProtection Features Resetting Latched ProtectionOvervoltage OvercurrentControl Connector OverpowerOvertemperature Reverse VoltageExternal Programming Input FaultMonitor Outputs Port On/Off Inspection Check Line Voltage Location and CoolingTurn-On Checkout Voltage Select Switches Line Voltage Fuse Agilent Part NoConnect The Power Cord Gpib Errors Input Errors Display DescriptionError Turn-On/SelftestController Connection Power TestGpib Connector Rear Panel Connectors and Switches Wire Size Strip backGpib Address AWGInput Binding Post Control Connector Pins 3 +S and -SIM and VM CommonTrigger Connector Sense SwitchApplication Connections Zero-Volt Loading Connections Maximum Wire Lengths to Limit Voltage DropsParallel Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Turning the Input On/Off Using The Function Keys∙ Short on Recommended Programming Sequence Setting the Mode of Operation Setting CC ValuesCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Enter/Output Statements Gpib AddressOutput EnterSending a Remote Command Remote Programming CommandsGetting Data Back Output 705 MEASCURR?Modes RangesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CR Mode Example CC Mode ExampleCV Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleHpsl Command Short Form Output 705TRAN on Output 705INPUT onRemote Programming Ranges Function Constant Current CCConstant Resistance CR 6060BConstant Voltage CV Transient OperationPage Equipment Characteristics Recommended Model Example ProgramsEquipment Required Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Program Listing PauseWait END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates