b.Press again and note that the display indicates "C:TLV 1.0000". Note that you can use the Input ENTRY keys to increment and decrement the transient current level. Operation is similar to that described above for the main current level.
Setting CR Values
The CR values are programmed by pressing the applicable FUNCTION keys and then setting the desired value using the ENTRY keys. The display identifies the selected function; for example, R:RNG identifies resistance range. See Appendix A for considerations regarding
Ranges
The resistance values can be programmed in a low, middle, or high range. The valid CR values that can be programmed are listed in Table
Table
Function
Key
Display
Range of Values
6060B6063B
Set Range |
| "R:RNG value" |
|
|
|
|
Low range |
|
| ³ 0 or £ 1 |
| ³ 0 and £ 24 | |
Middle range |
|
| > 1 or £ 1000 |
| > 24 or £ 24000 | |
High range |
|
| >1000 or | £ 10000 | > 24000 or £ 240000 | |
Set Main Level |
| "RES value" |
|
| (see Note 1) | |
Low range |
|
| 0.033 to 1.0000 |
| 0.200 to 24.000 | |
|
|
| ||||
Middle range |
|
| 1.0000 to 1000.0 |
| 24.000 to 24000 | |
High range |
|
| 10.000 to 10000 |
| 240.000 to 240000 | |
Set Slew Rate |
|
|
|
|
|
|
|
|
|
|
|
|
|
Low range | (shifted) | "V:SLW value" |
|
| (see Note 2) | |
Middle or High range |
| "C:SLW value" |
|
|
|
|
Set Transient Level |
| "R:TLV value" |
| (see Notes 1 and 3) | ||
Low range |
|
| 0.033 to 1.0000 |
| 0.200 to 24.000 | |
|
|
| ||||
Middle range |
|
| 1.0000 to 1000.0 |
| 24.000 to 24000 | |
High range |
|
| 10.000 to 10000 |
| 240.000 to 240000 | |
Notes: |
|
|
|
|
|
|
1.In the middle and high ranges, the resolution of the main level and the transient level degrades as higher values are
entered. The value of resistance displayed will be the closest one to the value entered. A similar effect will occur with
the and keys. Refer to Appendix A for considerations regarding high resistance applications.
2.In the low resistance range, the resistance slew rate is programmed in volts/microsecond instead of in ohms/microsecond. Whatever value is programmed for the voltage slew rate (see "Setting CV Values") is also used for resistance in the low range. In the middle and high ranges, the resistance slew rate is programmed in amps/microsecond. Whatever value is programmed for the current slew rate (see "Setting CC Values") is also used for resistance in either the middle or high ranges.
3.In the low range, the transient resistance level must be set to a higher value than the main resistance value. In the middle and high ranges, the transient resistance level must be set to a lower value than the main resistance value.
Changing the programming range can cause the present CR settings to be automatically adjusted to fit within the new range. For example, assume that you are programming the Agilent 6060B 300 Watt Electronic Load, the present range is 1 to 1 k ohms "R:RNG 1000.0", and the present settings are: