Agilent Technologies 5951-2826, 5951-2827 manual Controls and Indicators Description

Page 50

Item

3Electronic Load Status Annunicators

4 GPIB Status Annunicators

5 SYSTEM Keys

Table 4-1. Controls and Indicators (continued)

Description

CC-Indicates the Electronic Load is in the constant current (CC) mode.

Note that Figure 4-1 shows the Electronic Load is in the CC mode (CC annunciator is on).

CR-Indicates the Electronic Load is in the constant resistance (CR) mode.

CV-Indicates the Electronic Load is in the constant voltage (CV) mode.

Tran-Indicates that transient operation is enabled.

Unr-Indicates that the Electronic Load is unregulated (applies only in the CC mode and in the middle and high ranges of the CR mode).

Prot-Indicates when any protection features (CC, OV, OP, OT, etc.) are active.

Err-Indicates that remote programming error(s) have occurred.

Shift-Indicates that the shift key, bottom key (blue) in SYSTEM group, was pressed.

Rmt-Indicates that the Electronic Load is in the GPIB remote state. In the remote state, the only front panel key that will function is the Local key.

Addr-Indicates that the Electronic Load is addressed to talk or to listen over the GPIB.

SRQ-Indicates that the Electronic Load is requesting service over the GPIB; i.e., the service request line (SRQ) is active.

- Returns the Electronic Load from remote (computer) control to local (front panel) control.

- Displays the Electronic Load’s GPIB address. You can change the address using the numeric entry keys. You cannot query or change the address remotely (over the GPIB).

(shifted address key) - Displays error codes that resulted from remote programming.

- Used in conjunction with the ENTRY keys to recall the saved settings from the specified location (Recall 0 through Recall 7). Recall 7 recalls the factory default settings.

(shifted Recall key) - Used in conjunction with the ENTRY keys to save all of the present settings (mode, current, resistance, voltage, etc.) in the specified register (SAVE 0 thru SAVE 6). The settings in locations 1 thru 6 will be lost when ac power is cycled. However, SAVE 0 will cause the settings to be stored in non-volatile memory; and, the next time the Electronic Load is turned on, these settings will become the power on settings.

(blue shift Key) - Activates shifted key functions (e.g., Error, Save, Slew, etc.). The Shift annunciator goes on when this key is pressed.

50 Local Operation

Image 50
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Herstellerbescheinigung Safety SummarySymbol Description Manufacturer’s DeclarationPrinting History Table of Contents Calibration Local OperationRemote Operation Considerations for Operating in Constant Resistance Mode Options What’s In This ManualReader Path General InformationSafety Requirements SpecificationsPulsed Mode Regulation Constant Resistance Mode RangesModes Continuous Mode 6060B 6063BCurrent Readback Temperature Coefficient Supplemental Characteristics 6060B 6063BResolution Maximum Readback Capability BandwidthCurrent Slew Rate Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Rate # High Range Step Low Range Step Transition TimeModel 6063B Range Transient Current Level Resistance Slew RateOvershoot With unit off 6060B 6063B 10%With unit on WidthPage Remote Programming IntroductionFront Panel Description Operation OverviewConstant Current CC Mode Local/Remote ControlProgrammable Features Modes of OperationTriggered Current Level Software Current LimitImmediate Resistance Level Constant Resistance CR ModeSlew Rate Transient Resistance Level Constant Voltage CV ModeTriggered Resistance Level Transient Voltage Level Immediate Voltage LevelTriggered Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSets pulse width to 1 millisecond Selects the external trigger inputSelects pulsed transient operation Selects the external trigger input sourceToggling Triggering a preset levelTriggering a transient pulse Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesInput On/Off Saving and Recalling SettingsReading Remote Programming Errors Status ReportingOvervoltage Protection FeaturesResetting Latched Protection OvercurrentOvertemperature Control ConnectorOverpower Reverse VoltageExternal Programming Input FaultMonitor Outputs Port On/Off Inspection Check Line Voltage Location and CoolingTurn-On Checkout Voltage Select Switches Line Voltage Fuse Agilent Part NoError Connect The Power CordGpib Errors Input Errors Display Description Turn-On/SelftestController Connection Power TestGpib Connector Gpib Address Rear Panel Connectors and SwitchesWire Size Strip back AWGInput Binding Post Control Connector IM and VM Pins 3+S and -S CommonTrigger Connector Sense SwitchApplication Connections Zero-Volt Loading Connections Maximum Wire Lengths to Limit Voltage DropsParallel Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Turning the Input On/Off Using The Function Keys∙ Short on Recommended Programming Sequence CC Programming Ranges Function Setting the Mode of OperationSetting CC Values Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Output Enter/Output StatementsGpib Address EnterGetting Data Back Sending a Remote CommandRemote Programming Commands Output 705 MEASCURR?Transient levels ModesRanges Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CR Mode Example CC Mode ExampleCV Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleRemote Programming Ranges Function Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Constant Current CCConstant Voltage CV Constant Resistance CR6060B Transient OperationPage Equipment Required Equipment Characteristics Recommended ModelExample Programs Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Wait Program ListingPause END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates