Agilent Technologies 5951-2826, 5951-2827 manual Current Readback

Page 12

Table 1-1. Specifications (continued)

TRANSIENT CURRENT LEVEL

 

 

Ranges

 

 

Low Range:

0 to 6 A

0 to 1 A

High Range:

0 to 60 A

0 to 10 A

Accuracy

 

 

Low Range:

± 0.1% ± 80 mA

± 0.18% ± 13 mA

High Range:

± 0.1% ± 350 mA

± 0.18% ± 50 mA

 

6060B

6063B

 

TRANSIENT RESISTANCE LEVEL

 

 

Ranges

 

 

Low Range:

0.033 to 1 W

0.20 to 24 W

Middle Range:

1 to 1000 W

24 to 10,000 W

High Range:

10 to 10,000 W

240 to 50,000 W

Accuracy

 

 

Low Range:

±0.8% ± 10 m W

± 0.8% ± 200 m W

 

with ³ 6 A at input

with ³ 1 A at input

Middle Range:

± 0.3% ± 10 mS

± 0.3% ± 0.5 mS

 

with ³ 6 V at input

with ³ 24 V at input

High Range:

± 0.3% ± 10 mS

± 0.3% ± 0.4 mS

TRANSIENT VOLTAGE LEVEL

with ³ 6 V at input

with ³ 24 V at input

 

 

 

 

Range:

0 to 60 V

0 to 240 V

Accuracy:

0.1% ± 300 mV

± 0.15% ± 1.1 V

CURRENT READBACK

 

 

 

 

*Accuracy:

0.05% ± 65 mA

± 0.12% ± l0mA

*after 30 second wait.

 

 

VOLTAGE READBACK

 

 

 

 

Accuracy:

0.05% ± 45 mV

± 0.1%150 mV

POWER READBACK

 

 

 

 

Accuracy

±0.2% ± 4 W

±0.2% ± 3 W

PARD (20 Hz to 10 MHz noise)

 

 

 

 

Current:

4 mA rms/40 mA p-p

1 mA rms/10 mA p-p

Voltage:

6 mV rms

6 mV rms

 

 

 

12 General Information

Image 12
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Safety Summary Symbol DescriptionHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Local Operation Remote OperationCalibration Considerations for Operating in Constant Resistance Mode What’s In This Manual Reader PathOptions General InformationSafety Requirements SpecificationsRegulation Constant Resistance Mode Ranges Modes Continuous ModePulsed Mode 6060B 6063BCurrent Readback Supplemental Characteristics 6060B 6063B ResolutionTemperature Coefficient Maximum Readback Capability BandwidthSupplemental Characteristics Maximum VA Peak Inrush Current 2.5 a typicalCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeResistance Slew Rate OvershootModel 6063B Range Transient Current Level 6060B 6063B 10% With unit onWith unit off WidthPage Introduction Front Panel DescriptionRemote Programming Operation OverviewLocal/Remote Control Programmable FeaturesConstant Current CC Mode Modes of OperationTriggered Current Level Software Current LimitConstant Resistance CR Mode Slew RateImmediate Resistance Level Constant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Immediate Voltage Level Triggered Voltage LevelTransient Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSelects the external trigger input Selects pulsed transient operationSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a preset level Triggering a transient pulseToggling Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesSaving and Recalling Settings Reading Remote Programming ErrorsInput On/Off Status ReportingProtection Features Resetting Latched ProtectionOvervoltage OvercurrentControl Connector OverpowerOvertemperature Reverse VoltageFault Monitor OutputsExternal Programming Input Port On/Off Inspection Location and Cooling Turn-On CheckoutCheck Line Voltage Voltage Select Switches Line Voltage Fuse Agilent Part NoConnect The Power Cord Gpib Errors Input Errors Display DescriptionError Turn-On/SelftestPower Test Gpib ConnectorController Connection Rear Panel Connectors and Switches Wire Size Strip backGpib Address AWGInput Binding Post Control Connector Pins 3 +S and -SIM and VM CommonTrigger Connector Sense SwitchApplication Connections Maximum Wire Lengths to Limit Voltage Drops Parallel ConnectionsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Using The Function Keys ∙ Short onTurning the Input On/Off Recommended Programming Sequence Setting the Mode of Operation Setting CC ValuesCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Enter/Output Statements Gpib AddressOutput EnterSending a Remote Command Remote Programming CommandsGetting Data Back Output 705 MEASCURR?Modes RangesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CC Mode Example CV Mode ExampleCR Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleHpsl Command Short Form Output 705TRAN on Output 705INPUT onRemote Programming Ranges Function Constant Current CCConstant Resistance CR 6060BConstant Voltage CV Transient OperationPage Equipment Characteristics Recommended Model Example ProgramsEquipment Required Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Program Listing PauseWait END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates