Agilent Technologies 5951-2826 manual Output 705TRAN on Output 705INPUT on, Constant Current CC

Page 72

80 OUTPUT 705;"TRAN ON

90 OUTPUT 705;"INPUT ON"

 

.

 

.

 

.

 

.

 

200 OUTPUT 705;"*TRG"

 

210 END

Line 10:

Turns off Electronic Load input.

Line 20:

Selects the CR mode.

Line 30:

Selects the main resistance level to 10 ohms.

Line 40:

Sets the transient resistance level to 5 ohms. Remember in the 1 to 1k range, the transient resistance level

 

must be set to a lower level than the main resistance level.

Line 50:

Selects the GPIB as the trigger source.

Line 60:

Sets the current slew rate to the fastest rate. Remember that in the middle range, the resistance slew rate

 

is programmed in amps/second.

Line 70:

Selects pulsed transient operation and sets the pulse width to 1 millisecond.

Line 80:

Turns on the transient generator.

Line 90

Turns on Electronic Load input.

Line 100

 

to

Other commands are executed.

Line 190

 

Line 200: The *TRG command generates a 1 millisecond pulse at the Electronic load input.

Table 5-1. Remote Programming Ranges

Function

Constant Current (CC)

Set Range

Low Range

High Range

Set Main Level

Low Range

High Range

Set Slew Rate

Low Range

High Range

Set Transient Level

*Set Triggered Level

HPSL Command

(Short Form)

"CURR:RANG value"

’CURR value"

"CURR:SLEW value"

.

"CURR:TLEV value" "CURR:TRIG value"

Range of Values

6060B

6063B

³ 0 and £ 6 A

³ 0 and £1 A

> 6 and £ 60 A

> 1 and £ 10 A

0 to 6 A

0 to l A

0 to 60 A

0 to 10 A

100 to 500,000 A/s

1.7 to 83,000 A/s

1000 to 5,000,000 A/s

17 to 830,000 A/s

Same as CC main level

Same as CC main level

72 Remote Operation

Image 72
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Safety Summary Symbol DescriptionHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Local Operation Remote OperationCalibration Considerations for Operating in Constant Resistance Mode What’s In This Manual Reader PathOptions General InformationSafety Requirements SpecificationsRegulation Constant Resistance Mode Ranges Modes Continuous ModePulsed Mode 6060B 6063BCurrent Readback Supplemental Characteristics 6060B 6063B ResolutionTemperature Coefficient Maximum Readback Capability BandwidthSupplemental Characteristics Maximum VA Peak Inrush Current 2.5 a typicalCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeResistance Slew Rate OvershootModel 6063B Range Transient Current Level 6060B 6063B 10% With unit onWith unit off WidthPage Introduction Front Panel DescriptionRemote Programming Operation OverviewLocal/Remote Control Programmable FeaturesConstant Current CC Mode Modes of OperationTriggered Current Level Software Current LimitConstant Resistance CR Mode Slew RateImmediate Resistance Level Constant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Immediate Voltage Level Triggered Voltage LevelTransient Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSelects the external trigger input Selects pulsed transient operationSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a preset level Triggering a transient pulseToggling Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesSaving and Recalling Settings Reading Remote Programming ErrorsInput On/Off Status ReportingProtection Features Resetting Latched ProtectionOvervoltage OvercurrentControl Connector OverpowerOvertemperature Reverse VoltageFault Monitor OutputsExternal Programming Input Port On/Off Inspection Location and Cooling Turn-On CheckoutCheck Line Voltage Voltage Select Switches Line Voltage Fuse Agilent Part NoConnect The Power Cord Gpib Errors Input Errors Display DescriptionError Turn-On/SelftestPower Test Gpib ConnectorController Connection Rear Panel Connectors and Switches Wire Size Strip backGpib Address AWGInput Binding Post Control Connector Pins 3 +S and -SIM and VM CommonTrigger Connector Sense SwitchApplication Connections Maximum Wire Lengths to Limit Voltage Drops Parallel ConnectionsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Using The Function Keys ∙ Short onTurning the Input On/Off Recommended Programming Sequence Setting the Mode of Operation Setting CC ValuesCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Enter/Output Statements Gpib AddressOutput EnterSending a Remote Command Remote Programming CommandsGetting Data Back Output 705 MEASCURR?Modes RangesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CC Mode Example CV Mode ExampleCR Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleHpsl Command Short Form Output 705TRAN on Output 705INPUT onRemote Programming Ranges Function Constant Current CCConstant Resistance CR 6060BConstant Voltage CV Transient OperationPage Equipment Characteristics Recommended Model Example ProgramsEquipment Required Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Program Listing PauseWait END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates