Table 4-6. Factory Default Settings

Function

 

Setting

 

6060B

6063B

Input on/off

on

on

Short on/off

off

off

CURR level

0 A

0 A

CURR transient level

0 A

0 A

CURR slew rate

1 A/μs

0.83 A/μs

CURR range

60 A

10 A

*CURR protection level

61.2 A

10.2 A

*CURR protection delay

15 s

15 s

*CURR protection on/off

off

off

VOLT level

60 V

240 V

VOLT transient level

60 V

240 V

VOLT slew rate

5 V/μs

2 V/μs

RES level

1000 Ω

2400 Ω

RES transient level

1000 Ω

24000 Ω

RES range

1000 Ω

24000 Ω

Mode

CC

CC

Transient Operation

off

off

Frequency

1 kHz

1 kHz

Duty Cycle

50%

50%

**Transient mode

continuous

continuous

*Pulse width

0.5 ms

0.5 ms

*Port0 output signal

TTL logic 0

TTL logic 0

*Calibration mode

off

off

*Trigger source

hold

hold

*Can only be programmed remotely via the GPIB.

**Continuous transient operation is the only mode of transient operation available at the front panel. Pulsed, toggled, and continuous transient operating modes may be programmed remotely via the GPIB.

Changing "Wake-up" Settings

The "wake-up" settings are stored in register 0. At power-on, the Electronic Load will "wake-up" with these values set. When the Electronic Load is shipped from the factory, its "wake-up" values are the same as its factory default values (see Table 4-6).

You can change the "wake-up" values to whatever values you wish. You do this by setting them into the Electronic Load and then saving them in register 0 by pressing

(blue shift key)

(shifted Recall key)

.

 

 

When power is turned off and on, the Electronic Load will be set to the values you saved in register 0.

The Save 0 operation takes a few seconds to complete. Do not turn power off until the "SAVE 0 " message goes away indicating that the operation is complete. If you turn off power before completion, the Electronic Load’s non- volatile memory will be corrupted and the Electronic Load

will need to be recalibrated.

Local Operation 63

Image 63
Contents Operating Manual Agilent Part No Microfiche Part No Updated AprilCertification Safety Summary Manufacturer’s Declaration Safety SummarySymbol Description HerstellerbescheinigungPrinting History Table of Contents Local Operation Remote OperationCalibration Considerations for Operating in Constant Resistance Mode General Information What’s In This ManualReader Path OptionsSpecifications Safety Requirements6060B 6063B Regulation Constant Resistance Mode RangesModes Continuous Mode Pulsed ModeCurrent Readback Supplemental Characteristics 6060B 6063B ResolutionTemperature Coefficient Bandwidth Maximum Readback CapabilityRate # High Range Step Low Range Step Transition Time Supplemental CharacteristicsMaximum VA Peak Inrush Current 2.5 a typical Current Slew RateResistance Slew Rate OvershootModel 6063B Range Transient Current Level Width 6060B 6063B 10%With unit on With unit offPage Operation Overview IntroductionFront Panel Description Remote ProgrammingModes of Operation Local/Remote ControlProgrammable Features Constant Current CC ModeSoftware Current Limit Triggered Current LevelConstant Resistance CR Mode Slew RateImmediate Resistance Level Constant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Transient Operation Immediate Voltage LevelTriggered Voltage Level Transient Voltage LevelPulsed Transient Operation Hpsl Command DescriptionSelects the external trigger input source Selects the external trigger inputSelects pulsed transient operation Sets pulse width to 1 millisecondSlew Rate And Minimum Transition Time Triggering a preset levelTriggering a transient pulse TogglingRisetime Transition Limitation Input Current, Voltage, and Power MeasurementTransition Times and Slew Rates Short On/OffStatus Reporting Saving and Recalling SettingsReading Remote Programming Errors Input On/OffOvercurrent Protection FeaturesResetting Latched Protection OvervoltageReverse Voltage Control ConnectorOverpower OvertemperatureFault Monitor OutputsExternal Programming Input Port On/Off Inspection Location and Cooling Turn-On CheckoutCheck Line Voltage Line Voltage Fuse Agilent Part No Voltage Select SwitchesTurn-On/Selftest Connect The Power CordGpib Errors Input Errors Display Description ErrorPower Test Gpib ConnectorController Connection AWG Rear Panel Connectors and SwitchesWire Size Strip back Gpib AddressInput Binding Post Control Connector Common Pins 3+S and -S IM and VMSense Switch Trigger ConnectorApplication Connections Maximum Wire Lengths to Limit Voltage Drops Parallel ConnectionsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Local Operation Front Panel Controls and Indicators DescriptionControls and Indicators Description Function Keys Local Control Overview Using The Function Keys ∙ Short onTurning the Input On/Off Recommended Programming Sequence Key Display Setting the Mode of OperationSetting CC Values CC Programming Ranges FunctionExamples Setting CR Values Examples Setting CV Values Function Key Display Range of Values Continuous Pulse Train Programming RangesShorting The Input Using The System KeysDisplaying Error Codes Setting The Electronic Load’s Gpib AddressChanging Wake-up Settings Factory Default Settings Function 6060B 6063BRecalling the Factory Default Values Enter Enter/Output StatementsGpib Address OutputOutput 705 MEASCURR? Sending a Remote CommandRemote Programming Commands Getting Data BackSlew Rates ModesRanges Transient levelsRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CC Mode Example CV Mode ExampleCR Mode Example Pulsed Transient Operation Example Continuous Transient Operation ExampleConstant Current CC Hpsl Command Short FormOutput 705TRAN on Output 705INPUT on Remote Programming Ranges FunctionTransient Operation Constant Resistance CR6060B Constant Voltage CVPage Equipment Required for Calibration Equipment Characteristics Recommended ModelExample Programs Equipment RequiredCalibration Commands Example Program Calibration FlowchartsCalibration Information 6060B 6063B Supply Shunt Values SettingsCalibration Flowchart Calibration Flowchart Calibration Flowchart END if Program ListingPause WaitPrint Voltage Calibration Pause SubendElse Line ExplanationPage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates