+S and | Used to connect the remote sense leads to the power source. Pin 1 connects the + S signal and pin 2 |
| connects the - S signal. |
IM and VM | Used to monitor the Electronic Load’s input current and voltage. A 0 |
(pins 3 and 4) | appropriate pin indicates the |
| monitors voltage (VM). |
Common | Provides the common connection for the IM, VM, and external programming (Ext Prg) signals. This |
(pin 5) | common point is floating from ground at the potential of the - INPUT terminal. |
Ext Prg (pin 6) | Connects an external programming input. The CC and CV mode can be programmed with a 0 |
| V signal (ac or dc). This signal can act alone or can be summed with values programmed over the |
| GPIB. Thus, it is possible to have an ac signal applied at pin 6 upon a programmed dc level. |
Pin 7 | Not used. |
Flt (pin 8) | A |
| voltage condition occurs. This signal powers up in the inactive |
Port (pin 9) | A |
| programmed ON. This signal can be used to control an external device such as a relay for shorting the |
| Electronic Load’s input terminals or as a general purpose digital output port. This signal powers up in |
| the inactive |
Common | Provides the common connection for the Flt and Port signals. |
(pin 10) |
|
Replace the mating plug in the connector after you have finished making all wire connections. Replace the safety cover.
42 Installation