Program Listing (continued)

1030

!

1040

SUB Cal_res(@Ld,Chan,Res_rng,Res_hipt,Res_lopt,Flag)

1050

PRINT "RESISTANCE CALIBRATION, RANGE";Res_rng

1060

PRINT "Set power supply to calibration information table"

1070

PRINT "Press CONT when ready to continue"

1080

PAUSE

1090

OUTPUT @Ld;"CHAN";Chan

1100

OUTPUT @Ld;"MODE:RES"

1110

OUTPUT @Ld;"RES:RANG";Res_rng

1120

OUTPUT @Ld;"RES";Res_hipt

1130

INPUT "Enter voltage across input terminals in volts",Hipt_volt

1140

INPUT "Enter current through current shunt in amps”,Hipt_curr

1150

Hipt_res=Hipt_volt/Hipt_curr

1160

OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_res

1170

OUTPUT @Ld;"RES";Res_lopt

1180

INPUT "Enter voltage across input terminals in volts",Lopt_volt

1190

INPUT "Enter current through current shunt in amps",Lopt_curr

1200

Lopt_res=Lopt_volt/Lopt_curr

1210

OUTPUT @Ld;"CAL:LEV:LOW;Lopt_res

1220

PRINT "Test unit to verify resistance values"

1230

PRINT "Press CONT when ready to calibrate transient level"

1240

PAUSE

1250

IF Flag THEN

1260

OUTPUT @Ld;"RES";Res_hipt

1270

ELSE

1280

OUTPUT @Ld;"RES";Res_lopt

1290

END IF

1300

OUTPUT @Ld;"TRAN:STAT ON;MODE TOGG;:TRIG:SOUR BUS"

1310

IF Flag THEN

1320

OUTPUT @Ld;"RES:TLEV";Res_lopt

1330

ELSE

1340

OUTPUT @Ld;"RES:TLEV";Res_hipt

1350

END IF

1360

OUTPUT @Ld;"*TRG"

1370

INPUT "Enter voltage across input terminals in volts",Tran_volt

1380

INPUT "Enter current through current shunt in amps",Tran_curr

1390

Tran_res=Tran_volt/Tran_curr

1400

OUTPUT @Ld;"CAL:TLEV";Tran_res

1410

OUTPUT @Ld;"TRAN OFF"

1420

PRINT "Test unit to verify transient values are in spec"

1430

PRINT "Press CONT when ready to end program or calibrate next range"

1440

PAUSE

1450

SUBEND

84 Calibration

Image 84
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Safety Summary Symbol DescriptionHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Local Operation Remote OperationCalibration Considerations for Operating in Constant Resistance Mode What’s In This Manual Reader PathOptions General InformationSafety Requirements SpecificationsRegulation Constant Resistance Mode Ranges Modes Continuous ModePulsed Mode 6060B 6063BCurrent Readback Supplemental Characteristics 6060B 6063B ResolutionTemperature Coefficient Maximum Readback Capability BandwidthSupplemental Characteristics Maximum VA Peak Inrush Current 2.5 a typicalCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeResistance Slew Rate OvershootModel 6063B Range Transient Current Level 6060B 6063B 10% With unit onWith unit off WidthPage Introduction Front Panel DescriptionRemote Programming Operation OverviewLocal/Remote Control Programmable FeaturesConstant Current CC Mode Modes of OperationTriggered Current Level Software Current LimitConstant Resistance CR Mode Slew RateImmediate Resistance Level Constant Voltage CV Mode Triggered Resistance LevelTransient Resistance Level Immediate Voltage Level Triggered Voltage LevelTransient Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSelects the external trigger input Selects pulsed transient operationSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a preset level Triggering a transient pulseToggling Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesSaving and Recalling Settings Reading Remote Programming ErrorsInput On/Off Status ReportingProtection Features Resetting Latched ProtectionOvervoltage OvercurrentControl Connector OverpowerOvertemperature Reverse VoltageFault Monitor OutputsExternal Programming Input Port On/Off Inspection Location and Cooling Turn-On CheckoutCheck Line Voltage Voltage Select Switches Line Voltage Fuse Agilent Part NoConnect The Power Cord Gpib Errors Input Errors Display DescriptionError Turn-On/SelftestPower Test Gpib ConnectorController Connection Rear Panel Connectors and Switches Wire Size Strip backGpib Address AWGInput Binding Post Control Connector Pins 3 +S and -SIM and VM CommonTrigger Connector Sense SwitchApplication Connections Maximum Wire Lengths to Limit Voltage Drops Parallel ConnectionsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview Using The Function Keys ∙ Short onTurning the Input On/Off Recommended Programming Sequence Setting the Mode of Operation Setting CC ValuesCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Enter/Output Statements Gpib AddressOutput EnterSending a Remote Command Remote Programming CommandsGetting Data Back Output 705 MEASCURR?Modes RangesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CC Mode Example CV Mode ExampleCR Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleHpsl Command Short Form Output 705TRAN on Output 705INPUT onRemote Programming Ranges Function Constant Current CCConstant Resistance CR 6060BConstant Voltage CV Transient OperationPage Equipment Characteristics Recommended Model Example ProgramsEquipment Required Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Program Listing PauseWait END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates