Table
Voltage Slew Rate:
| 6060B (0 to 60V) | 6063B (0 to 240V) | Transition |
Rate # | Voltage Step | Voltage Step | Time* |
1 | 1 V/ms | 4 V/ms | 8.0 ms |
2 | 2.5 V/ms | 10 V/ms | 3.2 ms |
3 | 5 V/ms | 20 V/ms | 1.6 ms |
4 | 10 V/ms | 40 V/ms | 800 μs |
5 | 25 V/ms | 100 V/ms | 320 μs |
6 | 50 V/ms | 200 V/ms | 160 μs |
7 | 0.1 V/μs | 0.4 V/μs | 100 μs |
8 | 0.25 V/μs | 1 V/μs | 100 μs |
9 | 0.5 V/μs | 2 V/μs | 100 μs |
*Transition time is based on low capacitance current source.
Resistance Slew Rate |
|
|
|
|
Low Range: | Uses the value programmed for the voltage slew rate. |
| ||
Middle and High Ranges: | Uses the value programmed for the current slew rate. |
| ||
TRANSIENT CURRENT OVERSHOOT (When programmed from 0A): |
| |||
Model 6060B |
|
|
|
|
Range | Transient Current Level | Current Slew Rate | Overshoot* | |
60 A |
| All slew rates | 0 | |
| 3 A | I A/μs to 5 A/μs | 1% | |
| 3 A | I A/ms to 0.5 A/μs | 0 | |
6 A | 6 A |
| All slew rates | 0 |
| 3 A | 0.25 | A/μs and 0.5 A/μs | 1% |
| 3 A | 0.1 A/ms to 0.1 A μs | 0 | |
Model 6063B |
|
|
|
|
Range | Transient Current Level | Current Slew Rate | Overshoot* | |
10A |
| All slew rates | 0 | |
| 0.5 A | 0.17 | A/μs to 0.83 A/μs | 5% |
| 0.5 A | 0.17 | A/ms to 42 A/ms | 0 |
| 1 A |
| 0.83 A/μs | 1% |
| 1 A | 0.17 | A/ms to 0.17 A/μs | 0 |
1 A | 0.5 A |
| 83 A/ms | 4% |
| 0.5 A | 17 | A/s to 17 A/ms | 0 |
| 1 A |
| All slew rates | 0 |
*All overshoot values assume a total inductance of 1 μH, or less, in the load leads connected to the D.U.T. For Model 6060B, overshoot may be higher during first five seconds of programming if the unit has been operating at full current.