Agilent Technologies 5951-2826, 5951-2827 manual Resistance Slew Rate, Overshoot

Page 16

Table 1-2. Supplemental Characteristics (continued)

Voltage Slew Rate:

 

6060B (0 to 60V)

6063B (0 to 240V)

Transition

Rate #

Voltage Step

Voltage Step

Time*

1

1 V/ms

4 V/ms

8.0 ms

2

2.5 V/ms

10 V/ms

3.2 ms

3

5 V/ms

20 V/ms

1.6 ms

4

10 V/ms

40 V/ms

800 μs

5

25 V/ms

100 V/ms

320 μs

6

50 V/ms

200 V/ms

160 μs

7

0.1 V/μs

0.4 V/μs

100 μs

8

0.25 V/μs

1 V/μs

100 μs

9

0.5 V/μs

2 V/μs

100 μs

*Transition time is based on low capacitance current source.

Resistance Slew Rate

 

 

 

 

Low Range:

Uses the value programmed for the voltage slew rate.

 

Middle and High Ranges:

Uses the value programmed for the current slew rate.

 

TRANSIENT CURRENT OVERSHOOT (When programmed from 0A):

 

Model 6060B

 

 

 

 

Range

Transient Current Level

Current Slew Rate

Overshoot*

60 A

6-60 A

 

All slew rates

0

 

3 A

I A/μs to 5 A/μs

1%

 

3 A

I A/ms to 0.5 A/μs

0

6 A

6 A

 

All slew rates

0

 

3 A

0.25

A/μs and 0.5 A/μs

1%

 

3 A

0.1 A/ms to 0.1 A μs

0

Model 6063B

 

 

 

 

Range

Transient Current Level

Current Slew Rate

Overshoot*

10A

2-10 A

 

All slew rates

0

 

0.5 A

0.17

A/μs to 0.83 A/μs

5%

 

0.5 A

0.17

A/ms to 42 A/ms

0

 

1 A

 

0.83 A/μs

1%

 

1 A

0.17

A/ms to 0.17 A/μs

0

1 A

0.5 A

 

83 A/ms

4%

 

0.5 A

17

A/s to 17 A/ms

0

 

1 A

 

All slew rates

0

*All overshoot values assume a total inductance of 1 μH, or less, in the load leads connected to the D.U.T. For Model 6060B, overshoot may be higher during first five seconds of programming if the unit has been operating at full current.

16 General Information

Image 16
Contents Agilent Part No Microfiche Part No Updated April Operating ManualCertification Safety Summary Safety Summary Symbol DescriptionHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Remote Operation Local OperationCalibration Considerations for Operating in Constant Resistance Mode What’s In This Manual Reader PathOptions General InformationSafety Requirements SpecificationsRegulation Constant Resistance Mode Ranges Modes Continuous ModePulsed Mode 6060B 6063BCurrent Readback Resolution Supplemental Characteristics 6060B 6063BTemperature Coefficient Maximum Readback Capability BandwidthSupplemental Characteristics Maximum VA Peak Inrush Current 2.5 a typicalCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeOvershoot Resistance Slew RateModel 6063B Range Transient Current Level 6060B 6063B 10% With unit onWith unit off WidthPage Introduction Front Panel DescriptionRemote Programming Operation OverviewLocal/Remote Control Programmable FeaturesConstant Current CC Mode Modes of OperationTriggered Current Level Software Current LimitSlew Rate Constant Resistance CR ModeImmediate Resistance Level Triggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Immediate Voltage Level Triggered Voltage LevelTransient Voltage Level Transient OperationHpsl Command Description Pulsed Transient OperationSelects the external trigger input Selects pulsed transient operationSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a preset level Triggering a transient pulseToggling Slew Rate And Minimum Transition TimeInput Current, Voltage, and Power Measurement Risetime Transition LimitationShort On/Off Transition Times and Slew RatesSaving and Recalling Settings Reading Remote Programming ErrorsInput On/Off Status ReportingProtection Features Resetting Latched ProtectionOvervoltage OvercurrentControl Connector OverpowerOvertemperature Reverse VoltageMonitor Outputs FaultExternal Programming Input Port On/Off Inspection Turn-On Checkout Location and CoolingCheck Line Voltage Voltage Select Switches Line Voltage Fuse Agilent Part NoConnect The Power Cord Gpib Errors Input Errors Display DescriptionError Turn-On/SelftestGpib Connector Power TestController Connection Rear Panel Connectors and Switches Wire Size Strip backGpib Address AWGInput Binding Post Control Connector Pins 3 +S and -SIM and VM CommonTrigger Connector Sense SwitchApplication Connections Parallel Connections Maximum Wire Lengths to Limit Voltage DropsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Front Panel Controls and Indicators Description Local OperationControls and Indicators Description Function Keys Local Control Overview ∙ Short on Using The Function KeysTurning the Input On/Off Recommended Programming Sequence Setting the Mode of Operation Setting CC ValuesCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Continuous Pulse Train Programming Ranges Function Key Display Range of ValuesUsing The System Keys Shorting The InputSetting The Electronic Load’s Gpib Address Displaying Error CodesFactory Default Settings Function 6060B 6063B Changing Wake-up SettingsRecalling the Factory Default Values Enter/Output Statements Gpib AddressOutput EnterSending a Remote Command Remote Programming CommandsGetting Data Back Output 705 MEASCURR?Modes RangesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CV Mode Example CC Mode ExampleCR Mode Example Continuous Transient Operation Example Pulsed Transient Operation ExampleHpsl Command Short Form Output 705TRAN on Output 705INPUT onRemote Programming Ranges Function Constant Current CCConstant Resistance CR 6060BConstant Voltage CV Transient OperationPage Equipment Characteristics Recommended Model Example ProgramsEquipment Required Equipment Required for CalibrationCalibration Commands Calibration Flowcharts Example ProgramSupply Shunt Values Settings Calibration Information 6060B 6063BCalibration Flowchart Calibration Flowchart Calibration Flowchart Program Listing PauseWait END ifPause Subend Print Voltage CalibrationElse Explanation LinePage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates