Explanation

 

LINE 10-20

Specify select code, address, and channel (default 705, 1)

LINE 30

Turn calibration mode on

LINE 40-90

Assign variables for subprograms (see module calibration tables)

LINE 100

Store new constants in EEROM when calibration complete

LINE 110

Turn calibration mode off

LINE 140

Current calibration subroutine

LINE 200-220

Select channel, current mode, and range

LINE 230

Set high calibration point

LINE 240

If high current range, wait for internal current shunt to stabilize

LINE 260

Send measurement in amperes for high main calibration point

LINE 270

If high current range, send measurement in amperes for high readback cal point

LINE 280

Set low calibration point

LINE 300

Send measurement in amperes for low main calibration point

LINE 310

If high current range, send measurement in amperes for low readback cal point

LINE 350

Set low calibration point

LINE 360-370

Select transient toggle mode and GPIB trigger source

LINE 380-390

Turn transient mode on and set transient calibration point

LINE 400

Trigger transient level

LINE 410

If high current range, wait for internal current shunt to stabilize

LINE 430

Send measurement in amperes for high transient calibration point

LINE 440

Turn transient mode off

LINE 500

Voltage calibration subroutine

LINE 550-560

Select channel and voltage mode

LINE 570

Set high calibration point

LINE 590

Send measurement in volts for high main calibration point

LINE 600

Send measurement in volts for high readback calibration point

LINE 610

Set low calibration point

LINE 630

Send measurement in volts for low main calibration point

LINE 640

Send measurement in volts for low readback calibration point

LINE 680

Set low calibration point

LINE 690-700

Select transient toggle mode and GPIB trigger source

LINE 710-720

Turn transient mode on and set transient calibration point

LINE 730

Trigger transient level

LINE 750

Send measurement in volts for transient calibration point

LINE 760

Turn transient mode off

LINE 820

Resistance calibration subroutine

LINE 870-890

Select channel, resistance mode, and range

LINE 900

Set high calibration point

LINE 930-940

Calculate and send measurement in ohms for high main calibration point

LINE 950

Set low calibration point

LINE 980-990

Calculate and send measurement in ohms for low main calibration point

LINE 1030-1070

If middle and high range, set high calibration point; otherwise set low point

LINE 1080-1090

Select transient toggle mode and GPIB trigger source

LINE 1100

Turn transient mode on

LINE 1110-1150

If middle and high range, set lower transient point; otherwise set higher point

LINE 1160

Trigger transient level

LINE 1190-1200

Calculate and send measurement in ohms for transient calibration point

LINE 1210

Turn transient mode off

Calibration 85

Image 85
Contents Operating Manual Agilent Part No Microfiche Part No Updated AprilCertification Safety Summary Symbol Description Safety SummaryHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Table of Contents Remote Operation Local OperationCalibration Considerations for Operating in Constant Resistance Mode Reader Path What’s In This ManualOptions General InformationSpecifications Safety RequirementsModes Continuous Mode Regulation Constant Resistance Mode RangesPulsed Mode 6060B 6063BCurrent Readback Resolution Supplemental Characteristics 6060B 6063BTemperature Coefficient Bandwidth Maximum Readback CapabilityMaximum VA Peak Inrush Current 2.5 a typical Supplemental CharacteristicsCurrent Slew Rate Rate # High Range Step Low Range Step Transition TimeOvershoot Resistance Slew RateModel 6063B Range Transient Current Level With unit on 6060B 6063B 10%With unit off WidthPage Front Panel Description IntroductionRemote Programming Operation OverviewProgrammable Features Local/Remote ControlConstant Current CC Mode Modes of OperationSoftware Current Limit Triggered Current LevelSlew Rate Constant Resistance CR ModeImmediate Resistance Level Triggered Resistance Level Constant Voltage CV ModeTransient Resistance Level Triggered Voltage Level Immediate Voltage LevelTransient Voltage Level Transient OperationPulsed Transient Operation Hpsl Command DescriptionSelects pulsed transient operation Selects the external trigger inputSets pulse width to 1 millisecond Selects the external trigger input sourceTriggering a transient pulse Triggering a preset levelToggling Slew Rate And Minimum Transition TimeRisetime Transition Limitation Input Current, Voltage, and Power MeasurementTransition Times and Slew Rates Short On/OffReading Remote Programming Errors Saving and Recalling SettingsInput On/Off Status ReportingResetting Latched Protection Protection FeaturesOvervoltage OvercurrentOverpower Control ConnectorOvertemperature Reverse VoltageMonitor Outputs FaultExternal Programming Input Port On/Off Inspection Turn-On Checkout Location and CoolingCheck Line Voltage Line Voltage Fuse Agilent Part No Voltage Select SwitchesGpib Errors Input Errors Display Description Connect The Power CordError Turn-On/SelftestGpib Connector Power TestController Connection Wire Size Strip back Rear Panel Connectors and SwitchesGpib Address AWGInput Binding Post Control Connector +S and -S Pins 3IM and VM CommonSense Switch Trigger ConnectorApplication Connections Parallel Connections Maximum Wire Lengths to Limit Voltage DropsZero-Volt Loading Connections 11. Local Sensing 13. Parallel Operation Page Local Operation Front Panel Controls and Indicators DescriptionControls and Indicators Description Function Keys Local Control Overview ∙ Short on Using The Function KeysTurning the Input On/Off Recommended Programming Sequence Setting CC Values Setting the Mode of OperationCC Programming Ranges Function Key DisplayExamples Setting CR Values Examples Setting CV Values Function Key Display Range of Values Continuous Pulse Train Programming RangesShorting The Input Using The System KeysDisplaying Error Codes Setting The Electronic Load’s Gpib AddressChanging Wake-up Settings Factory Default Settings Function 6060B 6063BRecalling the Factory Default Values Gpib Address Enter/Output StatementsOutput EnterRemote Programming Commands Sending a Remote CommandGetting Data Back Output 705 MEASCURR?Ranges ModesTransient levels Slew RatesRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet CV Mode Example CC Mode ExampleCR Mode Example Pulsed Transient Operation Example Continuous Transient Operation ExampleOutput 705TRAN on Output 705INPUT on Hpsl Command Short FormRemote Programming Ranges Function Constant Current CC6060B Constant Resistance CRConstant Voltage CV Transient OperationPage Example Programs Equipment Characteristics Recommended ModelEquipment Required Equipment Required for CalibrationCalibration Commands Example Program Calibration FlowchartsCalibration Information 6060B 6063B Supply Shunt Values SettingsCalibration Flowchart Calibration Flowchart Calibration Flowchart Pause Program ListingWait END ifPrint Voltage Calibration Pause SubendElse Line ExplanationPage Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex 24, 25, 26 Index Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates