ESG Family Signal Generators |
Hardware missing; no such SIO address
The expected board is not installed. Report this error to the nearest
Agilent Technologies sales or service office.
A legal program command or query could not be executed because of a hardware problem in the device. The definition of what constitutes a hardware problem is completely
A legal program data element was parsed but could not be executed because the version of the data is incorrect to the device. This particular error is used when file or block data elements are recognized by the instrument, but cannot be executed for reasons of version incompatibility. For example, a
A legal program data element was parsed but could not be executed because the data format or structure is inappropriate. For example, when loading memory tables or when sending a SYSTem:SET parameter for an unknown instrument.
The measurement accuracy is questionable.
Data questionable; EEPROM copy of <filename>.
The EEPROM copy of a file has a correctable error. The system automatically performs the correction. A potential cause is a failing EEPROM. Report this error to the nearest Agilent Technologies sales or service office.
Data questionable; RAM copy of <filename>.
The
Possibly invalid data. A new reading was started but not completed since last access.
24 | Error Messages |