
201 to 702: | ESG Family Signal Generators |
621FPGA download failed; Demodulator board.
The download of the field programmable gate array code to the Demodulator board failed.
FPGA download failed; Opt UN7.
The download of the field programmable gate array code to the RAM on the A6 bit error rate tester board failed. Report this error to the nearest Agilent Technologies sales or service office.
622Sequence references self; <Sequence Name>.
The specified sequence is not allowed because it contains a reference to itself.
623Out of Lock; Demodulator VCO
The A/D 26 MHz clock is not correct. The problem may be with either the 10 MHz reference, or the 13 MHz GSM clock.
Out of Lock; Downconverter LO
There is a problem with the downconverter. The problem may be with either the 10 MHz reference, or the 13 MHz GSM clock.
700State Save Recall Error; Delete empty sequence <sequence_name>. Delete sequence ignored.
You have attempted to delete a sequence that is empty (all registers unused). This message is informational only. Typically this error is reported several times when the “Delete All Sequences” command is executed. If the file is displayed by catalog, delete explicitly.
State Save Recall Error; Delete sequence <sequence_name> ignored.
A STATE: file in a sequence that is being deleted could not be deleted due to the file not being found, data corruption, etc. If state file exists, delete explicitly using the memory catalog.
State Save Recall Error; Recall aborted. Unable to recall the state from
The state file was not readable, so the recall was aborted. If state file exists, delete explicitly using the memory catalog.
68 | Error Messages |