201 to 702: | ESG Family Signal Generators |
616Configuration Error; Attenuator configuration does not match actual installed attenuator type.
A mismatch was found between the configured and detected attenuator types. The instrument may not be fully functional. Report this error to the nearest Agilent Technologies sales or service office.
Configuration Error; Option reconfiguration failed.
As a part of option reconfiguration, the instrument reboots. If the reconfiguration failed, it is reported when the instrument completes the reboot. Try repeating the option reconfiguration. Report this error to the nearest Agilent Technologies sales or service office.
617Configuration error; Could not start license system.
The Software License System has failed to operate. The software option cannot be enabled. Report this error to the nearest Agilent Technologies sales or service office.
Configuration error; Data Generator Memory configuration does not match installed board.
The memory configuration for an option board does not match the known memory limits of the board. If this error has occurred as the result of a
Configuration error; Illegal combination of installed option boards.
An option board or board combination that is not supported by the installed firmware has been specified or detected. Be sure that you have the proper instrument and firmware to support the option board, and that you have specified a supported combination of options.
Configuration error; Incorrect demodulator board revision. An invalid Demodulator board is installed.
Configuration error; Incorrect downconverter board revision. An invalid Downconverter board is installed.
64 | Error Messages |