ESG Family Signal Generators |
System error; Cannot change manual point until list mode error condition cleared.
An error is keeping the sweep/list from being able to set the frequency and/or power. Until the problem is addressed, the manual point cannot be changed.
System error; File Not Found.
The specified file cannot be found. Verify that the file exists in the instrument file system.
System Error; FIR subsystem generated more coefficients than the hardware can handle.
The maximum number of coefficients for a given set of hardware has been exceeded. The filter must be created with fewer coefficients.
System error; Hardware error. Unable to apply Phase inversion!
There is an error with the UN8 option. Try turning CDMA2000 mode off and on again to reset the card. If the problem persists, report this error to the nearest Agilent Technologies sales or service office.
System error; RS232 buffer overflow: character lost.
The RS232 buffer has been exceeded. The most recent character has been dropped.
System error; Unable to determine which attenuator is installed.
An invalid attenuator identification code has been detected. Possible causes include a loose attenuator control cable. The instrument will likely not produce the proper output power levels. Report this error to the nearest Agilent Technologies sales or service office.
This is a generic
Error Messages | 15 |