-399 to -300: Device-Specific Errors

System error; Cannot change manual point until list mode error condition cleared.

An error is keeping the sweep/list from being able to set the frequency and/or power. Until the problem is addressed, the manual point cannot be changed.

System error; File Not Found.

The specified file cannot be found. Verify that the file exists in the instrument file system.

System Error; FIR subsystem generated more coefficients than the hardware can handle.

The maximum number of coefficients for a given set of hardware has been exceeded. The filter must be created with fewer coefficients.

System error; Hardware error. Unable to apply Phase inversion!

There is an error with the UN8 option. Try turning CDMA2000 mode off and on again to reset the card. If the problem persists, report this error to the nearest Agilent Technologies sales or service office.

System error; RS232 buffer overflow: character lost.

The RS232 buffer has been exceeded. The most recent character has been dropped.

System error; Unable to determine which attenuator is installed.

An invalid attenuator identification code has been detected. Possible causes include a loose attenuator control cable. The instrument will likely not produce the proper output power levels. Report this error to the nearest Agilent Technologies sales or service office.

-300Device-specific error

This is a generic device-dependent error for devices that cannot detect more specific errors. The code indicates only that a device-dependent error as defined in IEEE 488.2, 11.5.1.1.6 has occurred.

Error Messages

15

Image 17
Contents Supersedes February Copyright 1999-2001 Agilent Technologies Affix Label Here Part No. E4400-90333Contents Introduction Error Message Format Error Message Types No Error To -400 Query Errors Device has detected a failure during its self-test procedure To -300 Device-Specific ErrorsSelf-test for a particular power supply voltage has failed. Nearest Agilent Technologies sales or service officeBurst generator memory failed. Modulation data produced by 321Out of memory ARB communication failed Out of memory Could not copy current istate Out of memory Histogram display cannot function Out of memory Memory allocation error Out of memory Object Memory Area Instrument is still functional Lost. The meaning of this error is device-dependent320 315Protected user data saved by the *PUD command has been lost Non-volatile data saved by the *SAV? command has been lostAn error was detected in the device’s memory ValuesSystem error File Not Found To -200 Execution Errors 285 When parsing a downloaded program is device-specific284 IllegalDue to an error within the macro definition see Ieee 488.2 Command header, or contained invalid header syntaxAttempted. The definition of a math error is device-specific 276An illegal memory subsystem name was used after the @ filename of was specified. This is not a legal filenameFile name error Illegal filename character File name not found SYSTEMFCDMA2KCHANFILE Definition of what constitutes full media is device-specific 254Catalog. Afterwards, try again Try again252 Sales or service officeWhat constitutes missing media is device-specific 250Measurement accuracy is questionable Data corrupt or stale filename Algorithm. a potential cause is a failing backup battery During the first power up after downloading new firmware intoEeprom master copy if one exists. The system automatically 226Memory or related device-specific requirements 223To -200 Execution Errors Too much data User filter has too many symbols Data out of range Synthesizer Frequency out of bounds Data out of range value clipped to upper limit To -200 Execution Errors All digital modulation multicarriers must be the same format To -200 Execution Errors ESG Family Signal Generators Settings conflict Edge Bert requires Normal as Timeslot Type Settings conflict FM & PM not allowed Settings conflict GSM Bert requires Normal as Timeslot Type ESG Family Signal Generators Settings conflict I/Q & AM WB cannot be on at the same time ESG Family Signal Generators To -200 Execution Errors Used in RC3 and 4 for channel Used in RC1To -200 Execution Errors ESG Family Signal Generators Error For errors -221 throughManual mode is on and therefore the instrument cannot sweep 220212 Have described the error that is stalling list/sweepIgnored 211Execution Error Cannot update FlexDG with Even Second Delay To -100 Command Errors Specified filename exceeds the maximum length Detect a more specific error Device cannot detect a more specific error150 148Header contains more than twelve characters see Ieee 488.2 110Command header error Syntax error Invalid file name 201 to 702 Device-Specific ErrorsUser has attempted to write to a read-only memory subsystem State file is not readable and the recall was abortedInstrument’s output is unleveled Synthesizer is unlocked. Service may be neededInternal error report to factory 513 1 GHz Oscillator unlocked Calibration failure Dcfm DC overrange Reverse Power Protection RPP Tripped DSP FW download failed Demodulator board DSP times out Demodulator communication failed Option 300 DSP for Bert capability returns an error Configuration Error Option reconfiguration failed Configuration error Installed board not supported Configuration error OPTION, Invalid inconsistent license key Option UN7 hardware has an invalid revision of Fpga code Demodulator board has an invalid revision of DSP boot codeFpga download failed Demodulator board Modulation must have at least two states specified filter must have at least one coefficient specifiedBurst shape file is corrupted, or is not in the proper format file is either corrupt or is not a FIR file