Agilent Technologies E4400-90333 manual ESG Family Signal Generators

Page 47

ESG Family Signal Generators

-299 to -200: Execution Errors

Settings conflict; Walsh length invalid! Changed to default Walsh number.

The current Walsh length is invalid for the current data rate. For data rate 2X and above, doubling the rate halves the Walsh space. The channel Walsh number has been changed to the highest valid Walsh code.

Settings conflict; You may not set Forward Link values while in Reverse Link. Reset Link Direction to proceed.

Values or settings for forward link were entered while in reverse link. If forward link is the desired link direction, set the ESG to forward link first, and then re-enter the values or settings. However, if reverse link is the correct link direction, consult the Option 201 user’s and programming guide for the correct values or settings to enter.

Settings conflict; You may not set Reverse Link values while in Forward Link. Reset Link Direction to proceed.

Values or settings for reverse link were entered while in forward link. If reverse link is the desired link direction, set the ESG to reverse link first, and then re-enter the values or settings. However, if forward link is the correct link direction, consult the Option 201 user’s and programming guide for the correct values or settings to enter.

Settings conflict; You may only setup channels that are part of your current radio configuration.

An attempt was made to enter a value or setting for a channel that is not supported in the current radio configuration. The R-FCH and R-SCH are supported in all the radio configurations for reverse link. However, the values or settings allowed for these channels vary according to the radio configuration used. Refer to the Option 201 user’s and programming guide to determine the supported channels by radio configuration.

Error Messages

45

Image 47
Contents Supersedes February Copyright 1999-2001 Agilent Technologies Affix Label Here Part No. E4400-90333Contents Introduction Error Message Format Error Message Types No Error To -400 Query Errors Device has detected a failure during its self-test procedure To -300 Device-Specific Errors321 Nearest Agilent Technologies sales or service officeSelf-test for a particular power supply voltage has failed. Burst generator memory failed. Modulation data produced byOut of memory ARB communication failed Out of memory Could not copy current istate Out of memory Histogram display cannot function Out of memory Memory allocation error Out of memory Object Memory Area 315 Lost. The meaning of this error is device-dependentInstrument is still functional 320Values Non-volatile data saved by the *SAV? command has been lostProtected user data saved by the *PUD command has been lost An error was detected in the device’s memorySystem error File Not Found To -200 Execution Errors Illegal When parsing a downloaded program is device-specific285 284276 Command header, or contained invalid header syntaxDue to an error within the macro definition see Ieee 488.2 Attempted. The definition of a math error is device-specificAn illegal memory subsystem name was used after the @ filename of was specified. This is not a legal filenameFile name error Illegal filename character File name not found SYSTEMFCDMA2KCHANFILE Try again 254Definition of what constitutes full media is device-specific Catalog. Afterwards, try again250 Sales or service office252 What constitutes missing media is device-specificMeasurement accuracy is questionable Data corrupt or stale filename 226 During the first power up after downloading new firmware intoAlgorithm. a potential cause is a failing backup battery Eeprom master copy if one exists. The system automaticallyMemory or related device-specific requirements 223To -200 Execution Errors Too much data User filter has too many symbols Data out of range Synthesizer Frequency out of bounds Data out of range value clipped to upper limit To -200 Execution Errors All digital modulation multicarriers must be the same format To -200 Execution Errors ESG Family Signal Generators Settings conflict Edge Bert requires Normal as Timeslot Type Settings conflict FM & PM not allowed Settings conflict GSM Bert requires Normal as Timeslot Type ESG Family Signal Generators Settings conflict I/Q & AM WB cannot be on at the same time ESG Family Signal Generators To -200 Execution Errors Used in RC3 and 4 for channel Used in RC1To -200 Execution Errors ESG Family Signal Generators 220 For errors -221 throughError Manual mode is on and therefore the instrument cannot sweep211 Have described the error that is stalling list/sweep212 IgnoredExecution Error Cannot update FlexDG with Even Second Delay To -100 Command Errors Specified filename exceeds the maximum length 148 Device cannot detect a more specific errorDetect a more specific error 150Header contains more than twelve characters see Ieee 488.2 110Command header error Syntax error Invalid file name 201 to 702 Device-Specific ErrorsUser has attempted to write to a read-only memory subsystem State file is not readable and the recall was abortedInstrument’s output is unleveled Synthesizer is unlocked. Service may be neededInternal error report to factory 513 1 GHz Oscillator unlocked Calibration failure Dcfm DC overrange Reverse Power Protection RPP Tripped DSP FW download failed Demodulator board DSP times out Demodulator communication failed Option 300 DSP for Bert capability returns an error Configuration Error Option reconfiguration failed Configuration error Installed board not supported Configuration error OPTION, Invalid inconsistent license key Option UN7 hardware has an invalid revision of Fpga code Demodulator board has an invalid revision of DSP boot codeFpga download failed Demodulator board Modulation must have at least two states specified filter must have at least one coefficient specifiedBurst shape file is corrupted, or is not in the proper format file is either corrupt or is not a FIR file