201 to 702: | ESG Family Signal Generators |
Configuration error; Old bootrom detected. File system and instrument state
The firmware has detected that the instrument contains an old bootrom version. As a result the user file system, arb file system (if present), and instrument states have been
Configuration error; <OPTION>, Invalid (inconsistent) license key.
The License Key for the indicated software option is invalid. If the option has not been purchased, then enter a value of zero for its key to disable the option, otherwise enter the correct value from the License Key Certificate.
618RS232 times out; RS232 reset.
If further input is not received within the timeout time while a SCPI command is being processed, the command is aborted and the input buffer is cleared.
619DSP code is invalid; Bert board FPGA.
The host reads an unexpected FPGA version number from the Bert board. The FPGA version check is done by the DSP, so this error message is not generated if a DSP download Fail error message has already been generated. In this case, were do not know if the FPGA code stored in flash memory is correct.
DSP code is invalid; Bert DSP code is invalid.
The host reads an unexpected DSP version number from the demodulator board.
DSP code is invalid; Demodulator board FPGA.
The demodulator’s FPGA code is not the revision that the firmware expects.
DSP code is invalid; Demodulator DSP code is invalid.
The host reads and unexpected DSP version number from the demodulator board.
DSP code is invalid; Demodulator main DSP code is invalid. The demodulator board has an invalid revision of DSP firmware code.
66 | Error Messages |