ESG Family Signal Generators
Introduction
If an error condition occurs in the signal generator, it is reported to both the front panel display error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed separately.
NOTE | When there is an unviewed message in the front panel error queue, the ERR | |||
| annunciator appears on the signal generator’s display. | |||
|
|
|
| |
|
|
|
| |
| Characteristic | Front Panel Display | SCPI Remote Interface | |
|
| Error Queue | Error Queue | |
|
|
| ||
Capacity (#errors) | 30 | 30 | ||
|
|
|
| |
|
| Circular (rotating). | Linear, | |
Overflow Handling | Drops oldest error as new error comes in. | Replaces newest error | ||
| with: | |||
|
|
| ||
|
|
| overflow | |
|
|
|
| |
|
| Press: Utility > Error Info > View Next | Use SCPI query | |
Viewing Entries | SYSTem:ERRor? or | |||
(or Previous) Error Message | ||||
|
| STATus:QUEue? | ||
|
|
| ||
|
|
|
| |
|
|
| Power up | |
Clearing the Queue | Press: Utility > Error Info > Clear Error Queue(s) | Send a *CLS command | ||
|
|
| Read last item in the queue | |
|
|
| ||
Permanent Errors |
|
| ||
(errors that must be resolved. | ||||
For example: unlock, | cleared. | |||
| ||||
avalanched, and hi/lo) |
|
| ||
|
|
|
|