ESG Family Signal Generators |
Data corrupt or stale; <filename>.
The sequence file that you are attempting to run is the wrong size.
Data corrupt or stale; <NAME> differs offset 0x<VALUE>: NVRAM 0x<VALUE>, EEPROM 0x<VALUE>.
This message can occur only if full
Data corrupt or stale; EEPROM @ offset 0x<VALUE>: wrote 0x<VALUE>, read 0x<VALUE>.
After writing shadow RAM data to EEPROM, a memory comparison detected a difference. The EEPROM may not be retaining data properly. Report this error to the nearest Agilent Technologies sales or service office.
Data corrupt or stale; EEPROM copy of <filename>.
The EEPROM copy of a file is either corrupt or otherwise unusable. The system automatically updates the
Data corrupt or stale; Either the file is not a bit file, or the number of bits specified does not match with the number of data bytes in the file.
An error has occurred while downloading a user file to the Bit memory catalog. Either the user file is not a bit file, or the number of bits specified in the SCPI command does not match the number of bytes in the file. Ensure that the bit specifier matches the actual number of bits in the file.
Data corrupt or stale; file @ offset 0x<VALUE>: wrote 0x<VALUE>, retained 0x<VALUE>.
After writing shadow RAM data to EEPROM, a memory comparison detected a difference. The EEPROM may not be retaining data properly. Report this error to the nearest Agilent Technologies sales or service office.
Error Messages | 25 |