Calibration failure; DCFM DC overrange
The instrument was unable to perform a DCFM or DCΦM calibration due to the input signal being outside of the offset range that can be calibrated for.
Calibration failure; I/Q cal.
While executing the internal I/Q calibration, the attempt to allocate memory for the calibration failed.
Calibration failure; I/Q cal failed to allocate memory.
While executing the internal I/Q calibration, a failure occurred during the section indicated.
Calibration failure; ‘Marble Cal’, ‘Offset Cal’, ‘Other Cal’, ‘Gain Cal’, ‘BBG Cal’, ‘ARB Cal’
While executing the internal I/Q calibration, a failure occurred during the section indicated.
Calibration failure; Upgrade calibration failed. Data not stored.
The calibration stage of the instrument upgrade was not executed successfully. The calibration data has not been stored. The upgrade is not functional. Report this error to the nearest Agilent Technologies sales or service office.
518Arbitrary Waveform Generator unlocked; refer to manual.
The instrument’s dual arbitrary waveform generator is unlocked. This error will occur when the ARB Reference key is set to Ext and there is no external signal applied to the instrument. It may also occur if there is a problem with the A5 dual arbitrary waveform generator board.
519Uncalibrated signal; IQ Adjustments not calibrated. Refer to manual.
A calibration is required to use the IQ adjustments feature. The I/Q adjustments may be functional, but are not guaranteed to meet specifications until an I/Q calibration is performed. Refer to the user’s guide.