Agilent Technologies E4400-90333 manual File name error Illegal filename character

Page 22

-299 to -200: Execution Errors

ESG Family Signal Generators

File name error; Illegal filename character

An illegal character was used within a filename. \, :, @ and all non-printable ASCII characters are illegal in filenames.

File name error; Only one “:” is allowed.

Indicates that only one colon is allowed in any filename specification. The text before the colon is a user memory subsystem.

File name error; Only one “@” is allowed.

Indicates that only one @ is allowed in any filename specification. It specifies the memory subsystem that a user file actually resides in.

-256File name not found

A legal program command or query could not be executed because the file name on the device media could not be found. For example, an attempt was made to read or copy a nonexistent file. The definition of what constitutes a file not being found is device-specific.

File name not found; <filename>

The user has attempted to delete a file that does not exist, or an internal table editor column file has been lost.

File name not found; CDMA_DATA FILE

One of the internal working CDMA channel table files is missing.

File name not found; CDMA_WCODE FILE

One of the internal working CDMA channel table files is missing.

File name not found; CDMA_PNOFS FILE

One of the internal working CDMA channel table files is missing.

File name not found; CDMA_POW FILE

One of the internal working CDMA channel table files is missing.

File name not found; FCDMA2K_CHAN_FILE

An internal table editor column file has been lost. A new one will have to be created.

20

Error Messages

Image 22
Contents Affix Label Here Part No. E4400-90333 Supersedes February Copyright 1999-2001 Agilent TechnologiesContents Introduction Error Message Format Error Message Types No Error To -400 Query Errors To -300 Device-Specific Errors Device has detected a failure during its self-test procedureBurst generator memory failed. Modulation data produced by Nearest Agilent Technologies sales or service officeSelf-test for a particular power supply voltage has failed. 321Out of memory ARB communication failed Out of memory Could not copy current istate Out of memory Histogram display cannot function Out of memory Memory allocation error Out of memory Object Memory Area 320 Lost. The meaning of this error is device-dependentInstrument is still functional 315An error was detected in the device’s memory Non-volatile data saved by the *SAV? command has been lostProtected user data saved by the *PUD command has been lost ValuesSystem error File Not Found To -200 Execution Errors 284 When parsing a downloaded program is device-specific285 IllegalAttempted. The definition of a math error is device-specific Command header, or contained invalid header syntaxDue to an error within the macro definition see Ieee 488.2 276filename of was specified. This is not a legal filename An illegal memory subsystem name was used after the @File name error Illegal filename character File name not found SYSTEMFCDMA2KCHANFILE Catalog. Afterwards, try again 254Definition of what constitutes full media is device-specific Try againWhat constitutes missing media is device-specific Sales or service office252 250Measurement accuracy is questionable Data corrupt or stale filename Eeprom master copy if one exists. The system automatically During the first power up after downloading new firmware intoAlgorithm. a potential cause is a failing backup battery 226223 Memory or related device-specific requirementsTo -200 Execution Errors Too much data User filter has too many symbols Data out of range Synthesizer Frequency out of bounds Data out of range value clipped to upper limit To -200 Execution Errors All digital modulation multicarriers must be the same format To -200 Execution Errors ESG Family Signal Generators Settings conflict Edge Bert requires Normal as Timeslot Type Settings conflict FM & PM not allowed Settings conflict GSM Bert requires Normal as Timeslot Type ESG Family Signal Generators Settings conflict I/Q & AM WB cannot be on at the same time ESG Family Signal Generators To -200 Execution Errors Used in RC1 Used in RC3 and 4 for channelTo -200 Execution Errors ESG Family Signal Generators Manual mode is on and therefore the instrument cannot sweep For errors -221 throughError 220Ignored Have described the error that is stalling list/sweep212 211Execution Error Cannot update FlexDG with Even Second Delay To -100 Command Errors Specified filename exceeds the maximum length 150 Device cannot detect a more specific errorDetect a more specific error 148Header contains more than twelve characters see Ieee 488.2 110Command header error Syntax error 201 to 702 Device-Specific Errors Invalid file nameState file is not readable and the recall was aborted User has attempted to write to a read-only memory subsystemInternal error report to factory Synthesizer is unlocked. Service may be neededInstrument’s output is unleveled 513 1 GHz Oscillator unlocked Calibration failure Dcfm DC overrange Reverse Power Protection RPP Tripped DSP FW download failed Demodulator board DSP times out Demodulator communication failed Option 300 DSP for Bert capability returns an error Configuration Error Option reconfiguration failed Configuration error Installed board not supported Configuration error OPTION, Invalid inconsistent license key Demodulator board has an invalid revision of DSP boot code Option UN7 hardware has an invalid revision of Fpga codeFpga download failed Demodulator board filter must have at least one coefficient specified Modulation must have at least two states specifiedfile is either corrupt or is not a FIR file Burst shape file is corrupted, or is not in the proper format