Agilent Technologies E4400-90333 manual filter must have at least one coefficient specified

Page 71

ESG Family Signal Generators

201 to 702: Device-Specific Errors

State Save Recall Error; Recall data different from FW revision. No instrument state change.

An attempt was made to recall a state that was saved with an incompatible version of the instrument firmware. This typically occurs when a state file is copied from an instrument with a newer version of firmware to an instrument with an older version of firmware. Newer versions of instrument firmware can read older state files.

State Save Recall Error; Recall non-saved state register. Recall ignored.

A recall was attempted for a state register that is unused. If state file exists, delete explicitly using catalog.

State Save Recall Error; Recalled state has a bad checksum. No instrument state change.

The state file was corrupt or out-of-date, so the recall was ignored. If state file exists, delete explicitly using the memory catalog.

State Save Recall Error; The state file is from a different firmware revision that does not support comments.

An attempt was made to write a comment to a state file revision that does not support comments. Comments in saved state files are not supported by the A.01.00 and A.01.01 releases of the instrument firmware.

State Save Recall Error; Unable to read the secondary frame instrument state file. The secondary frame has been turned off.

The format that is currently active needed the secondary frame instrument state file to generate the secondary frame in pattern RAM, but the instrument state file was either not previously saved or it was deleted. You must create and save a secondary frame instrument state.

702Load/Store Error; Cannot create a filter with no coefficients.

A filter must have at least one coefficient specified.

Load/Store Error; Cannot create a Modulation with less than one bit per symbol.

A modulation must have at least two states specified.

Error Messages

69

Image 71
Contents Supersedes February Copyright 1999-2001 Agilent Technologies Affix Label Here Part No. E4400-90333Contents Introduction Error Message Format Error Message Types No Error To -400 Query Errors Device has detected a failure during its self-test procedure To -300 Device-Specific Errors321 Nearest Agilent Technologies sales or service officeSelf-test for a particular power supply voltage has failed. Burst generator memory failed. Modulation data produced byOut of memory ARB communication failed Out of memory Could not copy current istate Out of memory Histogram display cannot function Out of memory Memory allocation error Out of memory Object Memory Area 315 Lost. The meaning of this error is device-dependentInstrument is still functional 320Values Non-volatile data saved by the *SAV? command has been lostProtected user data saved by the *PUD command has been lost An error was detected in the device’s memorySystem error File Not Found To -200 Execution Errors Illegal When parsing a downloaded program is device-specific285 284276 Command header, or contained invalid header syntaxDue to an error within the macro definition see Ieee 488.2 Attempted. The definition of a math error is device-specificAn illegal memory subsystem name was used after the @ filename of was specified. This is not a legal filenameFile name error Illegal filename character File name not found SYSTEMFCDMA2KCHANFILE Try again 254Definition of what constitutes full media is device-specific Catalog. Afterwards, try again250 Sales or service office252 What constitutes missing media is device-specificMeasurement accuracy is questionable Data corrupt or stale filename 226 During the first power up after downloading new firmware intoAlgorithm. a potential cause is a failing backup battery Eeprom master copy if one exists. The system automaticallyMemory or related device-specific requirements 223To -200 Execution Errors Too much data User filter has too many symbols Data out of range Synthesizer Frequency out of bounds Data out of range value clipped to upper limit To -200 Execution Errors All digital modulation multicarriers must be the same format To -200 Execution Errors ESG Family Signal Generators Settings conflict Edge Bert requires Normal as Timeslot Type Settings conflict FM & PM not allowed Settings conflict GSM Bert requires Normal as Timeslot Type ESG Family Signal Generators Settings conflict I/Q & AM WB cannot be on at the same time ESG Family Signal Generators To -200 Execution Errors Used in RC3 and 4 for channel Used in RC1To -200 Execution Errors ESG Family Signal Generators 220 For errors -221 throughError Manual mode is on and therefore the instrument cannot sweep211 Have described the error that is stalling list/sweep212 IgnoredExecution Error Cannot update FlexDG with Even Second Delay To -100 Command Errors Specified filename exceeds the maximum length 148 Device cannot detect a more specific errorDetect a more specific error 150Header contains more than twelve characters see Ieee 488.2 110Command header error Syntax error Invalid file name 201 to 702 Device-Specific ErrorsUser has attempted to write to a read-only memory subsystem State file is not readable and the recall was abortedInstrument’s output is unleveled Synthesizer is unlocked. Service may be neededInternal error report to factory 513 1 GHz Oscillator unlocked Calibration failure Dcfm DC overrange Reverse Power Protection RPP Tripped DSP FW download failed Demodulator board DSP times out Demodulator communication failed Option 300 DSP for Bert capability returns an error Configuration Error Option reconfiguration failed Configuration error Installed board not supported Configuration error OPTION, Invalid inconsistent license key Option UN7 hardware has an invalid revision of Fpga code Demodulator board has an invalid revision of DSP boot codeFpga download failed Demodulator board Modulation must have at least two states specified filter must have at least one coefficient specifiedBurst shape file is corrupted, or is not in the proper format file is either corrupt or is not a FIR file