Agilent Technologies E4400-90333 150, Device cannot detect a more specific error, 148, 144, 141

Page 53

ESG Family Signal Generators-199 to -100: Command Errors

-150

String data error

 

This error, as well as errors -151 through -159, are generated when

 

parsing a string data element. This particular error message is used if the

 

device cannot detect a more specific error.

-148

Character data not allowed

 

A legal character data element was encountered where prohibited by the

 

device.

-144

Character data too long

 

The character data element contains more than twelve characters (see

 

IEEE 488.2, 7.7.1.4).

-141

Invalid character data

 

Either the character data element contains an invalid character or the

 

particular element received is not valid for the header.

-140

Character data error

 

This error, as well as errors -141 through -149, are generated when

 

parsing a character data element. This particular error message is used if

 

the device cannot detect a more specific error.

-138

Suffix not allowed

 

A suffix was encountered after a numeric element which does not allow

 

suffixes.

-134

Suffix too long

 

The suffix contained more than twelve characters (see IEEE 488.2,

 

7.7.3.4).

-131

Invalid suffix

 

The suffix does not follow the syntax described in IEEE 488.2, 7.7.3.2, or

 

the suffix is inappropriate for this device.

-130

Suffix error

 

This error, as well as errors -131 through -139, are generated when

 

parsing a suffix. This particular error message is used if the device cannot

 

detect a more specific error.

Error Messages

51

Image 53
Contents Supersedes February Copyright 1999-2001 Agilent Technologies Affix Label Here Part No. E4400-90333Contents Introduction Error Message Format Error Message Types No Error To -400 Query Errors Device has detected a failure during its self-test procedure To -300 Device-Specific ErrorsSelf-test for a particular power supply voltage has failed. Nearest Agilent Technologies sales or service officeBurst generator memory failed. Modulation data produced by 321Out of memory ARB communication failed Out of memory Could not copy current istate Out of memory Histogram display cannot function Out of memory Memory allocation error Out of memory Object Memory Area Instrument is still functional Lost. The meaning of this error is device-dependent320 315Protected user data saved by the *PUD command has been lost Non-volatile data saved by the *SAV? command has been lostAn error was detected in the device’s memory ValuesSystem error File Not Found To -200 Execution Errors 285 When parsing a downloaded program is device-specific284 IllegalDue to an error within the macro definition see Ieee 488.2 Command header, or contained invalid header syntaxAttempted. The definition of a math error is device-specific 276An illegal memory subsystem name was used after the @ filename of was specified. This is not a legal filenameFile name error Illegal filename character File name not found SYSTEMFCDMA2KCHANFILE Definition of what constitutes full media is device-specific 254Catalog. Afterwards, try again Try again252 Sales or service officeWhat constitutes missing media is device-specific 250Measurement accuracy is questionable Data corrupt or stale filename Algorithm. a potential cause is a failing backup battery During the first power up after downloading new firmware intoEeprom master copy if one exists. The system automatically 226Memory or related device-specific requirements 223To -200 Execution Errors Too much data User filter has too many symbols Data out of range Synthesizer Frequency out of bounds Data out of range value clipped to upper limit To -200 Execution Errors All digital modulation multicarriers must be the same format To -200 Execution Errors ESG Family Signal Generators Settings conflict Edge Bert requires Normal as Timeslot Type Settings conflict FM & PM not allowed Settings conflict GSM Bert requires Normal as Timeslot Type ESG Family Signal Generators Settings conflict I/Q & AM WB cannot be on at the same time ESG Family Signal Generators To -200 Execution Errors Used in RC3 and 4 for channel Used in RC1To -200 Execution Errors ESG Family Signal Generators Error For errors -221 throughManual mode is on and therefore the instrument cannot sweep 220212 Have described the error that is stalling list/sweepIgnored 211Execution Error Cannot update FlexDG with Even Second Delay To -100 Command Errors Specified filename exceeds the maximum length Detect a more specific error Device cannot detect a more specific error150 148Header contains more than twelve characters see Ieee 488.2 110Command header error Syntax error Invalid file name 201 to 702 Device-Specific ErrorsUser has attempted to write to a read-only memory subsystem State file is not readable and the recall was abortedInstrument’s output is unleveled Synthesizer is unlocked. Service may be neededInternal error report to factory 513 1 GHz Oscillator unlocked Calibration failure Dcfm DC overrange Reverse Power Protection RPP Tripped DSP FW download failed Demodulator board DSP times out Demodulator communication failed Option 300 DSP for Bert capability returns an error Configuration Error Option reconfiguration failed Configuration error Installed board not supported Configuration error OPTION, Invalid inconsistent license key Option UN7 hardware has an invalid revision of Fpga code Demodulator board has an invalid revision of DSP boot codeFpga download failed Demodulator board Modulation must have at least two states specified filter must have at least one coefficient specifiedBurst shape file is corrupted, or is not in the proper format file is either corrupt or is not a FIR file