| ESG Family Signal Generators |
| Parameter error |
| A program data element related error has occurred. This particular error |
| message is used if the device cannot detect a more specific error described |
| for errors |
| Parameter error; <modulation file name> |
| The specified file is not a modulation file type. |
| Arm deadlock |
| The arm source for the initiation of a measurement is set to GET and a |
| subsequent measurement query is received. The measurement cannot |
| begin until a GET is received, but the GET would cause an INTERRUPTED |
| error. |
| Trigger deadlock |
| A trigger source for the initiation of a measurement is set to GET and a |
| subsequent measurement query is received. The measurement cannot |
| begin until a GET is received, but the GET would cause an INTERRUPTED |
| error. |
| Trigger deadlock |
| TRIGger:SOURce BUS was selected when a MEASurement? query was |
| executed. No data is returned. |
| Init ignored |
| A request for a measurement initiation was ignored as another |
| measurement was already in progress. |
Init ignored; Cannot initiate sweep in manual mode.
The manual mode is on and therefore the instrument cannot sweep.
Init ignored; Sweep is already initiated.
The list/sweep is currently initiated and sweeping, therefore the command is not legal according to SCPI.
Init ignored; Sweep is already continuously initiated.
The list/sweep is continuously initiated and sweeping, therefore the command is not legal according to SCPI.
46 | Error Messages |