8.4 SETUP AND OPERATING INSTRUCTIONS

CAUTION

To prevent damage to the unit, do not program the output voltage and current to higher than the power supply rating.

8.4.1Setting up the power supply for 0-5V/0-10V Isolated Programming and Monitoring Perform the following procedure to configure the power supply:

1.Turn the power supply AC power switch to Off.

2.Connect a wire jumper between J1-8 and J1-12 (refer to Table 4-4).

3.Set the Setup switch SW1, positions 1 and 2 to the UP position.

4.Set SW1, position 3 to select the Programming Voltage Range: Down=0-5V, Up=0-10V.

5.Set SW1, position 4 to select the Monitoring Range: Down=0-5V, Up=0-10V.

6.Ensure that SW1, positions 7 and 8 are in the Down position.

7.Connect the programming sources to the mating plug of the Isolated Programming connector. Observe for correct polarity of the voltage source.

NOTE

J1-8 and J1-12 must be shorted together with a wire jumper.

8. Set the programming sources to the desired levels and turn the power supply ON.

8.4.2Setting up the power supply for 4-20mA Isolated Programming and Monitoring Perform the following procedure to configure the power supply:

1.Turn the power supply AC power switch to Off.

2.Connect a wire jumper between J1-8 and J1-12 (refer to Table 4-4).

3.Set the Setup switch SW1, positions 1 and 2 to the Up position.

4.Set SW1, position 3 to the Up position.

5.Set SW1, position 4 to the Up position.

6.Ensure that SW1 positions 1 and 2 to their Up position.

7.Connect the programming source to the mating plug of the Isolated Programming connector. Observe for correct polarity of the voltage source.

NOTE

J1-8 and J1-12 must be shorted together with a wire jumper.

8. Set the programming sources to the desired levels and turn the power supply ON.

NOTE

SW1 position 3 and 4 must be in the Up position for operation with 4-20mA Isolated Programming and Monitoring.

70

83-507-013 Rev. D

Image 78
Contents Genesystm 750W/1500W Page Page Genesys Manual Supplement Table of Contents Front and Rear Panel Controls and Connectors ………...……… Remote Analog PROGRAMMING……………………………………………… Isolated Analog Programming OPTION………………………………… Disclaimer Limitation of WarrantyWarranty Service Trademark InformationSafety Instructions Environmental Conditions SICHERHEITS-INSTALLATIONS Anweisungen Sicherheits-und Warnsymbole Anderungen and BauteileersatzUmweltbedingungen Introduction Features and optionsGeneral Information General DescriptionAnalog voltage programming and monitoring Multiple output power systemControl via the serial communication port Parallel operationAccessories Specifications ChapterVout Supplemental Characteristics Genesystm 750W & 1500W Power Supplies Outline Drawings Preparation for USE InstallationGeneral Initial InspectionsAC Source Requirements AC Input Power ConnectionLOCATON, Mounting and Cooling Rack Mount Slides optionalAC Input Wire Connection, 1500W models AC Input Connector, 1500W modelsAC Input Connector, 750W models AC Input CordTURN-ON Checkout Procedure GeneralPrior to Operation OVP Check Constant Voltage CheckConstant Current Check UVL CheckConnecting the Load Wire termination AWGNoise and Impedance Effects Inductive loadsMaking the load connections 6V to 60V Models Wire terminal lug 2 places 80V to 600V ModelsLoad wires connection to the output connector Connecting single loads, local sensing default Connecting single loads, remote sensingConnecting multiple loads, radial distribution method Multiple load connection with distribution terminals Grounding outputsLocal and Remote Sensing Sense wiringOutput Terminal Grounding Local sensing10.4 J2 sense connector technical information Remote sensingFront and Rear Panel Controls and Connectors Front Panel Controls and IndicatorsAuxiliary Function Rear Panel Connections and Controls Rear panel connections and controlsRear Panel SW1 Setup Switch Resetting the SW1 switch 1 SW1 position functionRear Panel J1 Programming and Monitorng Connector Making J1 connectionsIfcom SignalEnain LocalConstant Voltage Mode Constant Current ModeLocal Operation Over Voltage Protection OVP Setting the OVP levelResetting the OVP circuit Automatic CrossoverFoldback Protection Output SHUT-OFF SO Control VIA Rear Panel J1 ConnectorUnder Voltage Limit UVL Output ON/OFF ControlENABLE/DISABLE Control VIA Rear Panel J1 Connector CV/CC SignalPsok Signal Auto-restart mode Over Temperature Protecton OTPLast Setting Memory Safe-start modeSeries Operation Series connection for increased output voltageSeries connection for positive and negative output voltage Series connection for positive/negative output voltagesParallel Operation Master and Slave units default operation Advanced parallel configurationSetting the units as Master or Slave Connection to the loadTo release units from Slave mode Current display accuracyAST OFF *OUT OFF Unlocked front panel DAISY-CHAIN ConnectionFront Panel Locking Locked front panelRemote Analog Programming LOCAL/REMOTE Analog IndicationDown Remote voltage programming connectionResisitive Programming of Output Voltage and Current Limit Remote resistive programmingRemote Monitoring of Output Voltage and Current RS232 & RS485 Remote Control ConfigurationRear Panel RS232/RS485 Connector 6 RS232/RS485 port in Local modeFront panel in Remote mode MD Mode SRQ MD Mode Option Factory Installed MD Mode DescriptionMD Mode enable Serial communication mode MD Mode SRQ RetransmissionConnecting Power Supplies to RS232 or RS485 BUS Single power supplyCommunication Interface Protocol Error Messages Command SET DescriptionInitialization Control Commands ID Control CommandsOutput Control Commands FILTER? DVC?Filter OUT?Global Output Commands Disable SRQ retransmission MD Mode Option Required Single Byte CommandsEnable MD Mode MD Mode Option Required Enable SRQ retransmission MD Mode Option RequiredTest if MD Mode is Installed Addressed commands with response Read registersPrint Power On Time Addressed commands without response Acknowledge SRQ1010 None Disable retransmission of SRQs by GEN750W models GEN1500W models Current programming range STATUS, Error and SRQ Registers 10.6 Status Control CommandsConditional Registers Status RegistersService Request Enable and Event Registers BITFault Enable Register Fault Event RegisterStatus Enable Register Status Event RegisterPower supply set-up Serial Communication Test SET-UPCommunication Test 1 0-5V/0-10V Option PN IS510 SpecificationsIsolated Analog Programming Option PPM/CIsolated Programming & Monitoring Connector Setup and Operating Instructions Maintenance Fuse Rating OUP