TDK 750W, 1500W Constant Voltage Check, Constant Current Check, OVP Check, UVL Check

Page 25

2.Connect the unit to an AC source as described in Section 3.7.

3.Connect a DVM with appropriate cables for the rated voltage to the output terminals.

4.Turn the front panel AC power switch to On.

3.8.3Constant Voltage Check

1.Turn on the output by pressing the OUT pushbutton so the OUT LED illuminates.

2.Observe the power supply VOLT display and rotate the Voltage encoder. Ensure that the Output Voltage varies while the VOLT encoder is rotated. The minimum control range is from zero to the maximum rated output for the power supply model.

Compare the DVM reading with the front panel VOLT display to verify the accuracy of the VOLT display. Ensure that the front panel VOLT LED is On.

3.Turn Off the front panel AC power switch.

3.8.4Constant Current Check

1.Ensure that the front panel AC power switch is at the Off position and the DVM connected to the output terminals shows zero voltage.

2.Connect a DC shunt across the output terminals. Ensure that the shunt and the wire current rat- ings are higher than the power supply rating. Connect a DVM to the shunt.

3.Turn the front panel AC power switch to the On position,

4.Turn On the output by pressing OUT pushbutton so the OUT LED illuminates.

5.Observe the power supply CURRENT display and rotate the CURRENT encoder. Ensure that the Output Current varies while the CURRENT encoder is rotated. The minimum control range is from zero to the maximum rated output for the power supply model.

Compare the DVM reading with the front panel CURRENT display to verify the accuracy of the CURRENT display. Ensure that the front panel CURRENT LED is On.

6.Turn Off the front panel AC power switch.

7.Remove the shunt from the power supply output terminals.

3.8.5OVP Check

Refer to Section 5.3 for explanation of the OVP function prior to performing the procedure below.

1.Turn the front panel AC power switch to the On position and turn on the output by pressing OUT pushbutton.

2.Using the VOLT encoder, adjust the Output Voltage to approx. 10% of the unit voltage rating.

3.Momentarily press the OVP/UVL button so that the CURRENT display shows “OUP”. The VOLTAGE display will show the last setting of the OVP level.

4.Rotate the VOLT encoder CCW to adjust the OVP setting to 50% of the unit voltage rating.

5.Wait a few seconds until the VOLT display returns to show the Output Voltage.

6.Adjust the Output Voltage toward its maximum and check that the Output Voltage cannot be in- creased more than the OVP setting.

7.Adjust OVP limit to the maximum by repeating Step 3 and rotating the VOLT encoder C.W.

3.8.6UVL Check

Refer to Section 5.4 for explanation of the UVL function prior to performing the procedure below.

1.Press the OVP/UVL button TWICE so that the CURRENT display shows “UUL”. The VOLTAGE display will show the last setting of the UVL level.

2.Rotate the VOLT encoder to adjust the UVL level to approx. 10% of the unit voltage rating.

3.Wait a few seconds until the VOLT display returns to show the output voltage.

4.Adjust the output voltage toward its minimum and check that the output voltage cannot be de- creased below the UVL setting.

5.Adjust the UVL limit to the minimum by repeating Sep 1 and rotating the VOLT encoder CCW.

17

83-507-013 Rev. D

Image 25
Contents Genesystm 750W/1500W Page Page Genesys Manual Supplement Table of Contents Front and Rear Panel Controls and Connectors ………...……… Remote Analog PROGRAMMING……………………………………………… Isolated Analog Programming OPTION………………………………… Warranty Service Limitation of WarrantyDisclaimer Trademark InformationSafety Instructions Environmental Conditions SICHERHEITS-INSTALLATIONS Anweisungen Anderungen and Bauteileersatz Sicherheits-und WarnsymboleUmweltbedingungen General Information Features and optionsIntroduction General DescriptionControl via the serial communication port Multiple output power systemAnalog voltage programming and monitoring Parallel operationAccessories Chapter SpecificationsVout Supplemental Characteristics Genesystm 750W & 1500W Power Supplies Outline Drawings General InstallationPreparation for USE Initial InspectionsLOCATON, Mounting and Cooling AC Input Power ConnectionAC Source Requirements Rack Mount Slides optionalAC Input Connector, 750W models AC Input Connector, 1500W modelsAC Input Wire Connection, 1500W models AC Input CordGeneral TURN-ON Checkout ProcedurePrior to Operation Constant Current Check Constant Voltage CheckOVP Check UVL CheckConnecting the Load AWG Wire terminationInductive loads Noise and Impedance EffectsMaking the load connections 6V to 60V Models 80V to 600V Models Wire terminal lug 2 placesLoad wires connection to the output connector Connecting single loads, remote sensing Connecting single loads, local sensing defaultConnecting multiple loads, radial distribution method Grounding outputs Multiple load connection with distribution terminalsOutput Terminal Grounding Sense wiringLocal and Remote Sensing Local sensingRemote sensing 10.4 J2 sense connector technical informationFront Panel Controls and Indicators Front and Rear Panel Controls and ConnectorsAuxiliary Function Rear panel connections and controls Rear Panel Connections and ControlsRear Panel SW1 Setup Switch 1 SW1 position function Resetting the SW1 switchMaking J1 connections Rear Panel J1 Programming and Monitorng ConnectorEnain SignalIfcom LocalConstant Current Mode Constant Voltage ModeLocal Operation Resetting the OVP circuit Setting the OVP levelOver Voltage Protection OVP Automatic CrossoverUnder Voltage Limit UVL Output SHUT-OFF SO Control VIA Rear Panel J1 ConnectorFoldback Protection Output ON/OFF ControlCV/CC Signal ENABLE/DISABLE Control VIA Rear Panel J1 ConnectorPsok Signal Last Setting Memory Over Temperature Protecton OTPAuto-restart mode Safe-start modeSeries connection for increased output voltage Series OperationSeries connection for positive/negative output voltages Series connection for positive and negative output voltageParallel Operation Setting the units as Master or Slave Advanced parallel configurationMaster and Slave units default operation Connection to the loadCurrent display accuracy To release units from Slave modeAST OFF *OUT OFF Front Panel Locking DAISY-CHAIN ConnectionUnlocked front panel Locked front panelLOCAL/REMOTE Analog Indication Remote Analog ProgrammingRemote voltage programming connection DownRemote resistive programming Resisitive Programming of Output Voltage and Current LimitRemote Monitoring of Output Voltage and Current Configuration RS232 & RS485 Remote Control6 RS232/RS485 port in Local mode Rear Panel RS232/RS485 ConnectorFront panel in Remote mode MD Mode enable Serial communication mode MD Mode Option Factory Installed MD Mode DescriptionMD Mode SRQ MD Mode SRQ RetransmissionSingle power supply Connecting Power Supplies to RS232 or RS485 BUSCommunication Interface Protocol Command SET Description Error MessagesID Control Commands Initialization Control CommandsOutput Control Commands Filter DVC?FILTER? OUT?Global Output Commands Enable MD Mode MD Mode Option Required Single Byte CommandsDisable SRQ retransmission MD Mode Option Required Enable SRQ retransmission MD Mode Option RequiredPrint Power On Time Addressed commands with response Read registersTest if MD Mode is Installed Addressed commands without response Acknowledge SRQ1010 None Disable retransmission of SRQs by GEN750W models GEN1500W models Current programming range 10.6 Status Control Commands STATUS, Error and SRQ RegistersStatus Registers Conditional RegistersBIT Service Request Enable and Event RegistersFault Event Register Fault Enable RegisterStatus Event Register Status Enable RegisterSerial Communication Test SET-UP Power supply set-upCommunication Test Isolated Analog Programming Option Specifications1 0-5V/0-10V Option PN IS510 PPM/CIsolated Programming & Monitoring Connector Setup and Operating Instructions Maintenance OUP Fuse Rating