| Channel Trigger | Trigger modes and appendant registers | 
| 
 | 
 | 
Channel window trigger for entering signals
The analog input is continuously sampled with the selected sample rate. The upper and the lower level define a win- dow. Every time the signal enters the the window from the outside, a triggerevent will be detected.
| Register | Value | Direction | set to | Value | 
| SPC_TRIGGERMODE | 40000 | r/w | TM_CHANNEL | 20040 | 
| SPC_TRIGGERMODE0 | 40200 | r/w | TM_CHXWINENTER | 10040 | 
| 
 | 
 | 
 | 
 | 
 | 
| SPC_HIGHLEVEL0 | 42000 | r/w | Sets the window’s upper level relatively to the channel’s input range. | board dependant | 
| 
 | 
 | 
 | 
 | 
 | 
| SPC_LOWLEVEL0 | 42100 | r/w | Sets the window’s lower level relatively to the channel’s input range. | board dependant | 
Channel window trigger for leaving signals
The analog input is continuously sampled with the selected sample rate. The upper and the lower level define a win- dow. Every time the signal leaves the the window from the inside, a triggerevent will be detected.
| Register | Value | Direction | set to | Value | 
| SPC_TRIGGERMODE | 40000 | r/w | TM_CHANNEL | 20040 | 
| SPC_TRIGGERMODE0 | 40200 | r/w | TM_CHXWINLEAVE | 10050 | 
| SPC_HIGHLEVEL0 | 42000 | r/w | Sets the window’s upper level relatively to the channel’s input range. | board dependant | 
| 
 | 
 | 
 | 
 | 
 | 
| SPC_LOWLEVEL0 | 42100 | r/w | Sets the window’s lower level relatively to the channel’s input range. | board dependant | 
| 74 | MC.31xx Manual |