Quick Installation Guide
Rosemount 3095 MultiVariable
Canadian Standards Association (CSA)
CExplosion Proof for Class I, Division 1, Groups B, C, and D.
KIntrinsically Safe for Class I, Division 1, Groups A, B, C, and D. when installed in accordance with Rosemount drawing
For input parameters and installation see control drawing
European Certifications
FATEX Intrinsic Safety
Certificate Number: BAS98ATEX1359X II 1 G
Ex ia IIC T5 (Tamb =
Table 3. Connection Parameters (Power/Signal Terminals)
Ui = 30V
Ii = 200 mA
Pi = 1.0 W
Ci = 0.012 µF
Li = 0
Table 4. Temperature Sensor Connection Parameters
Uo = 30V
Io = 19 mA
Po = 140 mW
Ci = 0.002 µF
Li = 0
Table 5. Connection Parameters for Temperature Sensor Terminals
Co = 0.066 F | Gas Group IIC |
Co = 0.560 F | Gas Group IIB |
Co = 1.82 F | Gas Group IIA |
Lo = 96 mH | Gas Group IIC |
Lo = 365 mH | Gas Group IIB |
Lo = 696 mH | Gas Group IIA |
Lo/Ro = 247 H/ohm | Gas Group IIC |
Lo/Ro = 633 H/ohm | Gas Group IIB |
Lo/Ro= 633 H/ohm | Gas Group IIA |
Special Conditions for Safe Use
The 3095, when fitted with the transient terminal block (order code B), are not capable of withstanding the 500 volts insulation test required by EN
19