A&D HC-30Ki, HC-15Ki, HC-3Ki, HC-6Ki manual Cal 5001, Cal f 5000

Page 30

If you know the exact weight value, or

if you wish to use a different weight,

use the 0 Æ 9 and . 10-key pad to display the desired calibration weight.

(For example: Using 5 kg calibration weight that actually weighs 5.001 kg. Do not forget to enter a decimal point in this case.)

COUNT

888Cal10

WEIGHT

UNIT WEIGHT

Cal 0

5001.

3.Press the ENTER key.

The calibration weight stops blinking.

ENTER

COUNT

888Cal10

 

 

WEIGHT

 

UNIT WEIGHT

 

 

Cal 0

 

 

 

5000.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.Making sure that there is nothing on, or touching the weighing pan, press the ENTER key. When zero calibration is completed, the display will show “Cal f”.

If you don’t need span calibration, press the ON/OFF key to exit from the calibration procedure.

ENTER

Nothing on the

weighing pan!

 

COUNT

888Cal10

 

 

WEIGHT

 

UNIT WEIGHT

 

 

Cal f

 

 

 

5000.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.Place the calibration weight on the weighing pan and press the ENTER key.

When span calibration is completed, the display returns to step 1 showing the weight value for the calibration weight.

Remove the calibration weight.

If the calibration weight is not what it should be, an error will be displayed. Check if the weight is correct and try again.

6.Press the ON/OFF key to turn the scale off and re-attach the calibration switch cover.

(End of the calibration procedure.)

ENTER Place

Calibration

i ht

COUNT

888Cal02

WEIGHT

UNIT WEIGHT

5000.

If the scale will be moved to another place, set the gravity acceleration value before calibration. The value must be of the area where the calibration is to be done.

28

Image 30
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Features IntroductionIntroduction Unpacking Parts OverviewCalibration Setting Up Your ScaleBest Conditions For Weighing ON/OFFTotal 880000.880.00000Id-00 Automatic Last Unit Weight Used 888.8.888.00000 Front Panel Overview 880000.8 Turn The Power on and OFFBasic Operations Basic OperationsTare TareTo Start Counting Add Unit Weight By a Sample888.8.88.10 Add888.8.88.50 0000 Add0280 Add 888.8.88.20 0000 Add0112 Add 888.8.882.0 0000Key Unit Weight By Keyboard 888.8.88100000 Lo ut 0000 Id-00 Unit Weight By ID NumberReset Id-12888.8.881.0 Entering a Tare WeightUsing the Keyboard Tare Key 0615To Clear Tare 888.8.-554888.8-554 EitherStore Unit Weight Store Unit Weight by ID NumbersId-00 Clear idall Clear Id-00Clear Id-12 Clearing All ID Memories at OnceUnit Weight, Tare & Comparator Limits Memorized Key, the tare and/or comparator limitsM+ Memory Function Using the M+ MemoryAdding Using the M+ Key To Erase the Last M+ AdditionClearing the M+ Total Automatic M+ Accumulation ModeViewing the M+ Total M- Function Comparator Function 88888100 8f-00810 8f-05810 To Set the Comparator888.8.-198 888.8888.8.-102 Viewing Comparator Limits888Cal10 Calibration Procedure Using a Weight888Cal02 CalibrationCal f 5000 Cal 50015000 50009800 Gravity CompensationTo Change or View F-Function Settings Function Parameters01-01 00-0100-02 01-0202-01 Acai Operation & Min. Unit Weight01-05 02-0204-01 03-0103-02 04-0204-05 05-01 05-02 05-04 05-05Comparator 06-04 06-01 06-0206-03 RS-232C Data Output09-02 06-0509-01 09-X KeyAcai Function 8888882888888821 Acai Automatic OperationWeighing Pan Addition Range Acai Manual OperationPcs On Stable Acai888.off02 To Enable/Disable AWA FunctionOff Mode Setting AWA FunctionTarget 888.of100888.of150 IntOP-02 Battery Using OP-02 SLA BatteryCharging the Battery 88lo8batOP-03 RS-232C Serial Interface InstallationRS-232C Specifications Ascii1100 Data Output ModeData Output Mode f-06-01 Data to be Sent f-06-02Example of f-06-02settings Connecting the AD-8121 Printer / Mode 1 or ModePrint Operations Settings Baud Rate f-06-04Connecting the AD-8121 Printer / Mode Command ModeAD-8121 Printout Sample Command List Acknowledgment and Error Codes‰ Weight Data Data Format‰ Unit Weight ‰ Store the specified F-Function setting value Using UFC Universal Flex Coms Function Store Text Data into the Scale MemoryExamples of PF command and AD-8121 Printout Sample Data Format for the Scale DataOP-04 Installation Comparator Relay Output CircuitMaximum rating of the Relay Output Specifications ModelGravity Acceleration MAP Values of gravity at various locationsWorld map Blank