A&D HC-6Ki, HC-15Ki, HC-3Ki, HC-30Ki manual OP-04 Installation, Comparator Relay Output Circuit

Page 55

14. OP-04 RS-232C & RELAY OUTPUT

RS-232C interface and relay output for comparator results are obtained by installing OP-04.

‰The OP-04 unit includes an interface board, a connector plug (DIN type) and two screws. (M3x6 tapping type).

OP-04 Installation

‰See “13-1. Installation”.

‰OP-04 is installed to the same slot with the OP-03.

‰The RS-232C specifications are same as the OP-03 RS-232C interface and refer to “13. OP-03 RS-232C SERIAL INTERFACE”.

Comparator Relay Output Circuit

Pin connections

(View from outside)

Mating connector: JA+TCP0586 (Included in the OP-04)

1HI (relay output)

2Receive data (RS-232C)

3Transmit data (RS-232C)

4LO (relay output)

5Signal ground (RS-232C)

6OK (relay output)

7Data set ready (RS-232C)

8COM (relay common)

Inside of the HC-i

‰The HC-iis designated as DCE (Data Communication Equipment).

Maximum rating of the Relay Output

The maximum rating of the replay output is as follows.

‰ Maximum voltage:

50V DC

‰ Maximum current:

100mA DC

‰ Maximum ON resistance

30Ÿ

53

Image 55
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Introduction FeaturesIntroduction Parts Overview UnpackingON/OFF Setting Up Your ScaleBest Conditions For Weighing Calibration880000.880.00000 TotalId-00 888.8.888.0 Automatic Last Unit Weight Used0000 Front Panel Overview Basic Operations Turn The Power on and OFFBasic Operations 880000.8Tare TareTo Start Counting Add Unit Weight By a Sample888.8.88.10 Add0000 Add 888.8.88.500280 Add 0000 Add 888.8.88.200112 Add 0000 888.8.882.0Key 888.8.8810 Unit Weight By Keyboard0000 Lo ut Id-12 Unit Weight By ID NumberReset 0000 Id-000615 Entering a Tare WeightUsing the Keyboard Tare Key 888.8.881.0Either 888.8.-554888.8-554 To Clear TareStore Unit Weight by ID Numbers Store Unit WeightId-00 Clearing All ID Memories at Once Clear Id-00Clear Id-12 Clear idallKey, the tare and/or comparator limits Unit Weight, Tare & Comparator Limits MemorizedTo Erase the Last M+ Addition Using the M+ MemoryAdding Using the M+ Key M+ Memory FunctionM- Function Automatic M+ Accumulation ModeViewing the M+ Total Clearing the M+ TotalTo Set the Comparator 888881008f-00810 8f-05810 Comparator FunctionViewing Comparator Limits 888.8888.8.-102 888.8.-198Calibration Calibration Procedure Using a Weight888Cal02 888Cal105000 Cal 50015000 Cal f 5000Gravity Compensation 9800Function Parameters To Change or View F-Function Settings01-02 00-0100-02 01-0102-02 Acai Operation & Min. Unit Weight01-05 02-0104-02 03-0103-02 04-0105-01 05-02 05-04 05-05 04-05Comparator RS-232C Data Output 06-01 06-0206-03 06-0409-X Key 06-0509-01 09-02Acai Automatic Operation 8888882888888821 Acai FunctionStable Acai Acai Manual OperationPcs On Weighing Pan Addition RangeAWA Function To Enable/Disable AWA FunctionOff Mode Setting 888.off02Int 888.of100888.of150 TargetUsing OP-02 SLA Battery OP-02 Battery88lo8bat Charging the BatteryAscii InstallationRS-232C Specifications OP-03 RS-232C Serial InterfaceData to be Sent f-06-02 Data Output ModeData Output Mode f-06-01 1100Baud Rate f-06-04 Connecting the AD-8121 Printer / Mode 1 or ModePrint Operations Settings Example of f-06-02settingsCommand Mode Connecting the AD-8121 Printer / ModeAD-8121 Printout Sample Acknowledgment and Error Codes Command ListData Format ‰ Weight Data‰ Unit Weight ‰ Store the specified F-Function setting value Store Text Data into the Scale Memory Using UFC Universal Flex Coms FunctionData Format for the Scale Data Examples of PF command and AD-8121 Printout SampleComparator Relay Output Circuit OP-04 InstallationMaximum rating of the Relay Output Model SpecificationsValues of gravity at various locations Gravity Acceleration MAPWorld map Blank