A&D HC-6Ki, HC-15Ki, HC-3Ki Connecting the AD-8121 Printer / Mode 1 or Mode, Baud Rate f-06-04

Page 47

Baud Rate (f-06-04)

Select the baud rate according to the device to be connected.

‰2400 bps (f-06-04="0") Select 2400 bps to connect with an AD-8121.

‰4800 bps (f-06-04="1")

‰9600 bps (f-06-04="2")

13-4. Connecting the AD-8121 Printer / MODE 1 or MODE 2

‰When using the AD-8121 printer (MODE 1 or MODE 2), you will be able to get data: Number of data items, total, maximum, minimum, mean value, range of data (max. - min. data) and standard deviation.

‰When using the AD-8121 with MODE 2, set f-06-02 to print pcs (count) data only or weight data only.

‰To print date and time, use the AD-8121’s calendar / clock function and set f-06-02 to print pcs (count) data only or weight data only.

Print Operations Settings

Print By:

 

F-Function f-06-01

Printer MODE

HC

 

 

 

key

 

MODE 1

PRINT

0

 

 

 

 

 

Auto Print

 

2 or 3

MODE 1

Printer

 

 

key

 

MODE 2

DATA

1

 

 

 

 

 

 

 

Example of f-06-02settings

‰To print pcs (count) data only: set f-06-02at "0100"

‰To print weight data only: set f-06-02at "0010"

‰To print pcs (count) and weight data: set f-06-02at "0110"

‰To print pcs, weight and unit weight data: set f-06-02at "0111"

‰To print total data (accumulated by the M+ key), press the TOTAL count display shows the total, then press the PRINT key.

key so the

‰If you are using the AD-8121’s statistic functions, then set f-06-02at "01#0" (# = 0 or 1) for pcs (count) data or "0010" for weight data.

‰MODE 1 and 2 of the AD-8121 can not print ID numbers.

45

Image 47
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Introduction FeaturesIntroduction Parts Overview UnpackingON/OFF Setting Up Your ScaleBest Conditions For Weighing CalibrationId-00 Total880000.880.00000 0000 Automatic Last Unit Weight Used888.8.888.0 Front Panel Overview Basic Operations Turn The Power on and OFFBasic Operations 880000.8Tare TareTo Start Counting Add Unit Weight By a Sample888.8.88.10 Add0280 Add 888.8.88.500000 Add 0112 Add 888.8.88.200000 Add Key 888.8.882.00000 0000 Lo ut Unit Weight By Keyboard888.8.8810 Id-12 Unit Weight By ID NumberReset 0000 Id-000615 Entering a Tare WeightUsing the Keyboard Tare Key 888.8.881.0Either 888.8.-554888.8-554 To Clear TareId-00 Store Unit WeightStore Unit Weight by ID Numbers Clearing All ID Memories at Once Clear Id-00Clear Id-12 Clear idallKey, the tare and/or comparator limits Unit Weight, Tare & Comparator Limits MemorizedTo Erase the Last M+ Addition Using the M+ MemoryAdding Using the M+ Key M+ Memory FunctionM- Function Automatic M+ Accumulation ModeViewing the M+ Total Clearing the M+ TotalTo Set the Comparator 888881008f-00810 8f-05810 Comparator FunctionViewing Comparator Limits 888.8888.8.-102 888.8.-198Calibration Calibration Procedure Using a Weight888Cal02 888Cal105000 Cal 50015000 Cal f 5000Gravity Compensation 9800Function Parameters To Change or View F-Function Settings01-02 00-0100-02 01-0102-02 Acai Operation & Min. Unit Weight01-05 02-0104-02 03-0103-02 04-01Comparator 04-0505-01 05-02 05-04 05-05 RS-232C Data Output 06-01 06-0206-03 06-0409-X Key 06-0509-01 09-02Acai Automatic Operation 8888882888888821 Acai FunctionStable Acai Acai Manual OperationPcs On Weighing Pan Addition RangeAWA Function To Enable/Disable AWA FunctionOff Mode Setting 888.off02Int 888.of100888.of150 TargetUsing OP-02 SLA Battery OP-02 Battery88lo8bat Charging the BatteryAscii InstallationRS-232C Specifications OP-03 RS-232C Serial InterfaceData to be Sent f-06-02 Data Output ModeData Output Mode f-06-01 1100Baud Rate f-06-04 Connecting the AD-8121 Printer / Mode 1 or ModePrint Operations Settings Example of f-06-02settingsAD-8121 Printout Sample Connecting the AD-8121 Printer / ModeCommand Mode Acknowledgment and Error Codes Command ListData Format ‰ Weight Data‰ Unit Weight ‰ Store the specified F-Function setting value Store Text Data into the Scale Memory Using UFC Universal Flex Coms FunctionData Format for the Scale Data Examples of PF command and AD-8121 Printout SampleMaximum rating of the Relay Output OP-04 InstallationComparator Relay Output Circuit Model SpecificationsValues of gravity at various locations Gravity Acceleration MAPWorld map Blank