A&D HC-15Ki, HC-3Ki, HC-30Ki, HC-6Ki Function Parameters, To Change or View F-Function Settings

Page 32

9. F-FUNCTION PARAMETERS

9-1. To Change or View F-Function Settings

‰Start with the scale switched off.

1. Press and hold the

ZERO

key, then press

ZERO

the

ON/OFF

key.

 

 

 

+

ON

OFF

The count display will show “f-00” with “00” blinking. Then release the both keys.

2.Press the 0 Æ 9 keys to display the number of the F-Function section.

‰For example: the 5 key to enter into the F-Function F-05-X Comparator section.

8f-00810

8f-05810

3. Press the ENTER key.

The count display will show the

ENTER

F-Function and its present setting will

 

blink.

 

4.You may now either change the setting (Step

5)or move to the next F-Function (Step 6).

8f-05-01

0

5.Use the setting.

0

Æ

9

keys to change the 8f-05-01

The C key clears the input setting if you press the wrong key and want to re-enter.

1

If you make a mistake and want to escape without saving any changes made after the last time the ENTER key was pressed – press the ON/OFF key to exit.

After the ENTER key is pressed, the data is entered.

6. Press the ENTER key to save any

ENTER

changes and/or move to next.

7.When finished: press the ON/OFF key to exit. Then, press it to turn the display back ON. New settings will operate as set.

8f-05-02

0

30

Image 32
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Introduction FeaturesIntroduction Unpacking Parts OverviewSetting Up Your Scale Best Conditions For WeighingCalibration ON/OFFId-00 Total880000.880.00000 0000 Automatic Last Unit Weight Used888.8.888.0 Front Panel Overview Turn The Power on and OFF Basic Operations880000.8 Basic OperationsTare TareTo Start Counting Unit Weight By a Sample 888.8.88.10Add Add0280 Add 888.8.88.500000 Add 0112 Add 888.8.88.200000 Add Key 888.8.882.00000 0000 Lo ut Unit Weight By Keyboard888.8.8810 Unit Weight By ID Number Reset0000 Id-00 Id-12Entering a Tare Weight Using the Keyboard Tare Key888.8.881.0 0615888.8.-554 888.8-554To Clear Tare EitherId-00 Store Unit WeightStore Unit Weight by ID Numbers Clear Id-00 Clear Id-12Clear idall Clearing All ID Memories at OnceUnit Weight, Tare & Comparator Limits Memorized Key, the tare and/or comparator limitsUsing the M+ Memory Adding Using the M+ KeyM+ Memory Function To Erase the Last M+ AdditionAutomatic M+ Accumulation Mode Viewing the M+ TotalClearing the M+ Total M- Function88888100 8f-00810 8f-05810Comparator Function To Set the Comparator888.8 888.8.-102888.8.-198 Viewing Comparator Limits Calibration Procedure Using a Weight 888Cal02 888Cal10 CalibrationCal 5001 5000Cal f 5000 50009800 Gravity CompensationTo Change or View F-Function Settings Function Parameters00-01 00-0201-01 01-02Acai Operation & Min. Unit Weight 01-0502-01 02-0203-01 03-0204-01 04-02Comparator 04-0505-01 05-02 05-04 05-05 06-01 06-02 06-0306-04 RS-232C Data Output06-05 09-0109-02 09-X Key88888828 88888821Acai Function Acai Automatic OperationAcai Manual Operation Pcs OnWeighing Pan Addition Range Stable AcaiTo Enable/Disable AWA Function Off Mode Setting888.off02 AWA Function888.of100 888.of150Target IntOP-02 Battery Using OP-02 SLA BatteryCharging the Battery 88lo8batInstallation RS-232C SpecificationsOP-03 RS-232C Serial Interface AsciiData Output Mode Data Output Mode f-06-011100 Data to be Sent f-06-02Connecting the AD-8121 Printer / Mode 1 or Mode Print Operations SettingsExample of f-06-02settings Baud Rate f-06-04AD-8121 Printout Sample Connecting the AD-8121 Printer / ModeCommand Mode Command List Acknowledgment and Error Codes‰ Weight Data Data Format‰ Unit Weight ‰ Store the specified F-Function setting value Using UFC Universal Flex Coms Function Store Text Data into the Scale MemoryExamples of PF command and AD-8121 Printout Sample Data Format for the Scale DataMaximum rating of the Relay Output OP-04 InstallationComparator Relay Output Circuit Specifications ModelGravity Acceleration MAP Values of gravity at various locationsWorld map Blank