1-4. Setting Up Your Scale

1.Place the scale on a suitable weighing surface (see “Best Conditions For Weighing” below) and turn the adjustable feet until the spirit level shows that the scale is level.

2.Plug in the AC adapter. The AC input requirements could be l00, 120, 220, 230 or 240 Volts (50/60Hz) depending on the area in the world so please check that the adapter is correct. Earth the chassis if you think static electricity may be a problem.

3. Press the ON/OFF key to turn the power ON. All the display symbols are displayed.

AC adapter jack

Earth terminal

Then the display turns off once, and zero will be shown with the ZERO mark.

4. Pressing the

ON/OFF

key again, and the power will switch OFF.

‰Auto-power off function

It is possible to have the power automatically switched OFF. If zero is displayed for approximately 5 minutes. See “9-2. F-Functions” and set the F-Function f-04-05 at “1” to enable the function.

5.Switch the power ON at least half an hour before use so that the scale can warm up.

Best Conditions For Weighing

‰The Scale must be level (check the spirit level on the scale).

‰Best operating temperature is between 20°C~25°C / 68°F~77°F at about 50%~60% Relative Humidity. There shouldn’t be large temperature fluctuations.

‰The weighing room should be kept clean and dry.

‰The weighing table must be of a solid construction.

‰Corners of rooms are best as they are less prone to vibrations.

‰Don’t install the scale near heaters or air conditioners.

‰Don’t install the scale in direct sunshine.

‰Try to ensure a stable AC power supply when using an adapter.

‰Keep equipment containing magnets away from the scale.

‰Warm up the scale more than 30 minutes before use.

‰Ground the scale chassis for electrostatic discharge if the weighing conditions warrant.

Calibration

Calibration of the HC-iis required when the scale is initially installed. Please see “8. CALIBRATION” for more calibration information.

5

Image 7
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Introduction FeaturesIntroduction Parts Overview UnpackingON/OFF Setting Up Your ScaleBest Conditions For Weighing Calibration880000.880.00000 TotalId-00 888.8.888.0 Automatic Last Unit Weight Used0000 Front Panel Overview Basic Operations Turn The Power on and OFFBasic Operations 880000.8Tare TareTo Start Counting Add Unit Weight By a Sample888.8.88.10 Add0000 Add 888.8.88.500280 Add 0000 Add 888.8.88.200112 Add 0000 888.8.882.0Key 888.8.8810 Unit Weight By Keyboard0000 Lo ut Id-12 Unit Weight By ID NumberReset 0000 Id-000615 Entering a Tare WeightUsing the Keyboard Tare Key 888.8.881.0Either 888.8.-554888.8-554 To Clear TareStore Unit Weight by ID Numbers Store Unit WeightId-00 Clearing All ID Memories at Once Clear Id-00Clear Id-12 Clear idallKey, the tare and/or comparator limits Unit Weight, Tare & Comparator Limits MemorizedTo Erase the Last M+ Addition Using the M+ MemoryAdding Using the M+ Key M+ Memory FunctionM- Function Automatic M+ Accumulation ModeViewing the M+ Total Clearing the M+ TotalTo Set the Comparator 888881008f-00810 8f-05810 Comparator FunctionViewing Comparator Limits 888.8888.8.-102 888.8.-198Calibration Calibration Procedure Using a Weight888Cal02 888Cal105000 Cal 50015000 Cal f 5000Gravity Compensation 9800Function Parameters To Change or View F-Function Settings01-02 00-0100-02 01-0102-02 Acai Operation & Min. Unit Weight01-05 02-0104-02 03-0103-02 04-0105-01 05-02 05-04 05-05 04-05Comparator RS-232C Data Output 06-01 06-0206-03 06-0409-X Key 06-0509-01 09-02Acai Automatic Operation 8888882888888821 Acai FunctionStable Acai Acai Manual OperationPcs On Weighing Pan Addition RangeAWA Function To Enable/Disable AWA FunctionOff Mode Setting 888.off02Int 888.of100888.of150 TargetUsing OP-02 SLA Battery OP-02 Battery88lo8bat Charging the BatteryAscii InstallationRS-232C Specifications OP-03 RS-232C Serial InterfaceData to be Sent f-06-02 Data Output ModeData Output Mode f-06-01 1100Baud Rate f-06-04 Connecting the AD-8121 Printer / Mode 1 or ModePrint Operations Settings Example of f-06-02settingsCommand Mode Connecting the AD-8121 Printer / ModeAD-8121 Printout Sample Acknowledgment and Error Codes Command ListData Format ‰ Weight Data‰ Unit Weight ‰ Store the specified F-Function setting value Store Text Data into the Scale Memory Using UFC Universal Flex Coms FunctionData Format for the Scale Data Examples of PF command and AD-8121 Printout SampleComparator Relay Output Circuit OP-04 InstallationMaximum rating of the Relay Output Model SpecificationsValues of gravity at various locations Gravity Acceleration MAPWorld map Blank