A&D HC-6Ki, HC-15Ki, HC-3Ki, HC-30Ki manual 9800, Gravity Compensation

Page 31
888Cal10
COUNT

8-2. Gravity Compensation

When the scale is first used or has been moved to different place, it should be calibrated using a calibration weight.

But if the calibration weight cannot be prepared, the gravity acceleration correction will compensate the scale. Change the gravity acceleration value of the scale to the value of the area where it will be used. Refer to the gravity acceleration map appended to the end of this manual.

1.In the Step 1 above, press TARE key

The display flashes the gravity acceleration value stored in the scale.

WEIGHT

g

UNIT WEIGHT

9798.

2.Use the 0 Æ 9 10-key pad to

display the desired gravity acceleration value.

(Example of the value 9.800 m/s2.)

3.Press the ENTER key.

The scale will store the new value.

If necessary to calibrate the scale using a weight, go to Step 2 of the previous section.

COUNT

888Cal10

WEIGHT

UNIT WEIGHT

g

9800.

COUNT

ENTER 888Cal02

WEIGHT

UNIT WEIGHT

0000.

4.Press the ON/OFF key to turn the scale off and re-attach the calibration switch cover. (End of the calibration procedure.)

29

Image 31
Contents Counting Scale This manual and Marks Contents AWA Audible Weighing Assist Introduction FeaturesIntroduction Parts Overview UnpackingON/OFF Setting Up Your ScaleBest Conditions For Weighing Calibration880000.880.00000 TotalId-00 888.8.888.0 Automatic Last Unit Weight Used0000 Front Panel Overview Basic Operations Turn The Power on and OFFBasic Operations 880000.8Tare TareTo Start Counting Add Unit Weight By a Sample888.8.88.10 Add0000 Add 888.8.88.500280 Add 0000 Add 888.8.88.200112 Add 0000 888.8.882.0Key 888.8.8810 Unit Weight By Keyboard0000 Lo ut Id-12 Unit Weight By ID NumberReset 0000 Id-000615 Entering a Tare WeightUsing the Keyboard Tare Key 888.8.881.0Either 888.8.-554888.8-554 To Clear TareStore Unit Weight by ID Numbers Store Unit WeightId-00 Clearing All ID Memories at Once Clear Id-00Clear Id-12 Clear idallKey, the tare and/or comparator limits Unit Weight, Tare & Comparator Limits MemorizedTo Erase the Last M+ Addition Using the M+ MemoryAdding Using the M+ Key M+ Memory FunctionM- Function Automatic M+ Accumulation ModeViewing the M+ Total Clearing the M+ TotalTo Set the Comparator 888881008f-00810 8f-05810 Comparator Function Viewing Comparator Limits 888.8 888.8.-102 888.8.-198Calibration Calibration Procedure Using a Weight888Cal02 888Cal105000 Cal 50015000 Cal f 5000Gravity Compensation 9800Function Parameters To Change or View F-Function Settings01-02 00-0100-02 01-0102-02 Acai Operation & Min. Unit Weight01-05 02-0104-02 03-0103-02 04-0105-01 05-02 05-04 05-05 04-05Comparator RS-232C Data Output 06-01 06-0206-03 06-0409-X Key 06-0509-01 09-02Acai Automatic Operation 8888882888888821 Acai FunctionStable Acai Acai Manual OperationPcs On Weighing Pan Addition RangeAWA Function To Enable/Disable AWA FunctionOff Mode Setting 888.off02Int 888.of100888.of150 TargetUsing OP-02 SLA Battery OP-02 Battery88lo8bat Charging the BatteryAscii InstallationRS-232C Specifications OP-03 RS-232C Serial InterfaceData to be Sent f-06-02 Data Output ModeData Output Mode f-06-01 1100Baud Rate f-06-04 Connecting the AD-8121 Printer / Mode 1 or ModePrint Operations Settings Example of f-06-02settingsCommand Mode Connecting the AD-8121 Printer / ModeAD-8121 Printout Sample Acknowledgment and Error Codes Command ListData Format ‰ Weight Data‰ Unit Weight ‰ Store the specified F-Function setting value Store Text Data into the Scale Memory Using UFC Universal Flex Coms FunctionData Format for the Scale Data Examples of PF command and AD-8121 Printout SampleComparator Relay Output Circuit OP-04 InstallationMaximum rating of the Relay Output Model SpecificationsValues of gravity at various locations Gravity Acceleration MAPWorld map Blank