A&D EJ-6100, EJ-440, EJ-3000, EJ-4100, EJ-120, EJ-2000, EJ-1500 manual Units, Conversion table

Page 10

5-3. Units

The most common unit of weight used around the world is the gram, but there is often a need to shift to alternative units specific to the country where the balance is used or to select modes such as counting or percent.

The units and the order they appear in the display are as follows:

Ounce (avoir)

Pound

Metric carat

Percent mode

 

Counting mode

Momme

Pennyweight

Gram

Grain

Density Newton (specific gravity)

Among the units, those available for the user have been set at the factory before shipping.

The unit can be selected in the function setting mode. The order of the units available is the same as above.

‰Note

It is possible to store only the units that will be actually used from the units available. It is also possible to specify the display unit that will be shown first when the power is switched ON. For details, see “8-4. Storing weighing units”.

Conversion table

Units

Name

Conversion to gram

oz

Ounce (avoir)

28.349523125 g

lb

Pound (UK)

453.59237 g

ozt

Troy ounce

31.1034768 g

ct

Metric carat

0.2 g

mom

momme

3.75 g

dwt

Pennyweight

1.55517384 g

GN

Grain (UK)

0.06479891 g

t

tola

11.6638038 g

tl

tael (Hong Kong general, Singapore)

37.7994 g

tl

tael (Hong Kong jewelry)

37.4290 g

tl

tael (Taiwan)

37.5 g

‰“Newton” is the value calculated by “(g value) x (9.80665 m/s2) / 1000”.

‰The unit “t (tola)” and three kinds of “tl (tael)” are for special versions only. One of them can be selected and installed at the factory.

8

Image 10
Contents 1WMPD4001917 This manual and Marks Compliance with FCC rulesContents Maintenance Introduction UnpackingNET indicator Part Names and FunctionsSetting up your balance Setting UPPower source Breeze break Installing batteriesStorage Operation Turn the power on and OFFLCD backlight Units Conversion tableSelecting a weighing unit NETBasic operation Counting mode using the Acai function Counting mode pcsSelecting the counting mode Storing the sample unit100 0 % Percent mode %100 1000. %Comparator Setting exampleSelecting a comparison mode FuncCp lo 000000 Cp Hi 000000001234 001230Cal Calibration using a weightCalibration 060000Cal 50000 GlpCal 97985 Gravity acceleration correctionCp Hi Entering the function setting modeFunctions Key operationACai Func PoffEnd Cp Hi End id UnitStoring weighing units Function list Smpl Func ACaiUmin LdinEJ-02 USB Interface EJ-03 InstallationOptions EJ-03 RS-232C serial interfaceInterface specifications EIA RS-232CData output mode Data formatCommand mode Command listEJ-11 Breeze break EJ-07 / EJ-08 InstallationEJ-07 / EJ-08 Underhook EJ-12 Carrying caseEJ-13 Density Determination Kit EJ-13 InstallationDensity specific gravity measurement Change the function table 000Example of density measurement 2000 1751 8801End id func Setting the ID numberID Number and GLP Display character setData format for calibration report Output exampleData format for calibration test report End Block Calibration test using a weightTitle Title block and End block20000 20000 EndOverload error MaintenanceError codes Range over notice Unit weight error Sample quantity noticeLb0 Hb0 Error1Cal e -Cal e Hb1 Lb1Specifications EJ seriesOther weighing units Dimensions OptionsGravity Acceleration MAP Values of gravity at various locationsWorld map