A&D EJ-1500, EJ-6100, EJ-440 Functions, Cp Hi, Key operation, Entering the function setting mode

Page 19

8. FUNCTIONS

8-1. Key operation

Cancels the operation and turns off the balance.

Item key

Selects a class and an item.

In the weighing mode, press and hold the key to enter the function setting mode.

Enter key

Proceeds to the selected class.

Stores the settings per class and goes to the next class.

Parameter key

Selects a parameter.

Cancel key

Cancels the operation, and goes to the next class or returns to the weighing mode.

8-2. Entering the function setting mode

In the weighing mode, press and hold the

Press and hold

SAMPLE

key to enter the function setting mode

and display

func . Each time the SAMPLE

 

key is pressed, the class appears one after another.

Once the class is selected, the set items are available for selection. (See “Function list”.)

func

Each pressing switches the class

Cp Hi

Set using the relevant keys

17

Image 19
Contents 1WMPD4001917 Compliance with FCC rules This manual and MarksContents Maintenance Unpacking IntroductionPart Names and Functions NET indicatorSetting up your balance Setting UPPower source Breeze break Installing batteriesStorage Operation Turn the power on and OFFLCD backlight Conversion table UnitsSelecting a weighing unit NETBasic operation Storing the sample unit Counting mode pcsSelecting the counting mode Counting mode using the Acai function1000. % Percent mode %100 100 0 %Func Setting exampleSelecting a comparison mode Comparator001230 Cp Hi 000000001234 Cp lo 000000060000 Calibration using a weightCalibration CalGlp Cal 50000Gravity acceleration correction Cal 97985Key operation Entering the function setting modeFunctions Cp HiACai Func PoffEnd Cp Hi End id UnitStoring weighing units Function list Ldin Func ACaiUmin SmplEJ-03 RS-232C serial interface EJ-03 InstallationOptions EJ-02 USB InterfaceEIA RS-232C Interface specificationsData format Data output modeCommand list Command modeEJ-12 Carrying case EJ-07 / EJ-08 InstallationEJ-07 / EJ-08 Underhook EJ-11 Breeze breakEJ-13 Density Determination Kit EJ-13 InstallationDensity specific gravity measurement Change the function table 000Example of density measurement 1751 8801 2000Display character set Setting the ID numberID Number and GLP End id funcData format for calibration report Output exampleData format for calibration test report Title block and End block Calibration test using a weightTitle End Block20000 End 20000Range over notice Unit weight error Sample quantity notice MaintenanceError codes Overload errorHb1 Lb1 Error1Cal e -Cal e Lb0 Hb0EJ series SpecificationsOther weighing units Options DimensionsValues of gravity at various locations Gravity Acceleration MAPWorld map