Data format
S | T | , | + | 0 | 0 | 0 |
| 0 | 0 | . | 0 | 0 |
|
|
|
|
|
|
|
| g | CR | LF |
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Header |
|
|
| Data |
|
|
| Unit Terminator | |||||||||||||||||
|
|
|
|
|
|
|
|
|
| ||||||||||||||||
|
|
|
|
| Separator |
|
| (“ |
| ” shows a space.) | |||||||||||||||
|
|
|
|
| |||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
There are four types of headers:
ST : Stable weighing data (including % data)
QT : Stable counting data
US : Unstable weighing data (including count and %)
OL : Out of weighing range (Over)
The data is normally 9 digits including a decimal point and a sign.
There are 14 types of units:
|
|
|
| g | : Weighing data “gram” |
|
|
| |||
|
|
| P C | : Counting data “pcs” | |
|
|
| |||
|
|
|
| % | : Percentage data “%” |
|
| ||||
|
|
| o z | : Weighing data “decimal ounce” | |
|
|
| |||
|
|
| l b | : Weighing data “decimal pound” | |
|
|
| |||
o z t | : Weighing data “troy ounce” | ||||
|
|
| c t | : Weighing data “carat” | |
|
|
| |||
mom | : Weighing data “momme” | ||||
d w t | : Weighing data “penny weight” | ||||
|
| GN | : Weighing data “grain” | ||
|
| ||||
|
|
|
| N | : Force data “Newton” |
|
|
|
| ||
|
|
| t l | : Weighing data “tael” | |
|
|
| |||
|
|
|
| t | : Weighing data “tola” |
|
|
|
| ||
|
| DS | : Calculated density (specific gravity) value | ||
|
|
The terminator is always CRLF.
Example of output data:
Weighing data “gram” Counting data Percentage data
Out of range “gram” (+) Out of range “pcs”
S | T | , | + | 0 | 0 | 1 | 2 | 3 | 4 | . | 5 |
|
|
|
|
|
| g | CR | LF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Q | T | , | + | 0 | 0 | 0 | 1 | 2 | 3 | 4 | 5 |
|
|
|
| P |
| C | CR | LF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
S | T | , | + | 0 | 0 | 0 | 1 | 2 | 3 | . | 4 |
|
|
|
|
|
| % | CR | LF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
O | L | , | + | 9 | 9 | 9 | 9 | 9 | 9 | . | 9 |
|
|
|
|
|
| g | CR | LF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
O | L | , | - | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 |
|
|
|
| P |
| C | CR | LF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Data output mode
Stream mode
Set the function “prt 0”.
The balance outputs the current display data. The
The balance does not output data while it is in the setting mode.