2.
CC 0
is displayed.
CC 0
3.If necessary to change the weight value to check, press the change the weight value using the following keys.
SAMPLE
key and
SAMPLE |
| key | To select the digit blinking to change. | |
|
|
| key | To set the value of the digit selected. |
|
| |||
| key | To store the value and return to step 2. | ||
4. At step 2, press the PRINT key. The zero point is weighed and the weighed value is displayed for a few seconds.
With nothing on the pan
The weighed data is shown with “g”.
5.Place the displayed weight on the pan and press the
PRINT key to weigh it. The weighed value is displayed for a few seconds.
The weighed data is shown with “g”.
6. | end | appears. |
CC 0
00.
20000.
Displayed weight
20000.
20000.
end
7.glp
output.
8.end press the mode.
is displayed and calibration test report is
appears again. Remove the weight and MODE key to return to the weighing
glp GLP output
end
Returns to
the weighing mode
33