A&D EJ-120, EJ-6100, EJ-440, EJ-3000, EJ-200, EJ-4100 manual Cal e -Cal e, Lb0 Hb0, Hb1 Lb1, Error1

Page 37

50 -

100 -

CAL errors

Cal e -Cal e

Battery error

lb0

Hb0

AC adapter error

Hb1

lb1

Stability error

error1

the unit weight.

Note: Pressing the PRINT key without adding samples may reduce counting accuracy. Starting from the 100 samples, 100 - may be displayed when the sample weight is light. This is for your notice and press the PRINT key without adding any samples.

When “aCai 0” (ACAI disabled) or “Umin 2” is set, this notice is not shown.

Warning to indicate that calibration has been canceled because the calibration weight is too heavy.

Warning to indicate that calibration has been canceled because the calibration weight is too light.

Check the weighing pan and the calibration weight.

To return to the weighing mode, press the UNITS key.

Warning to show that the batteries are exhausted. Replace them with new one or use the AC adapter.

Warning to show that the battery voltage is too high. Check the battery.

Warning to show that the output voltage of an AC adapter is too high. Check if the AC adapter is correct.

Warning to show that the output voltage of an AC adapter is too low. Check if the AC adapter is correct.

Warning to indicate that the weight value is not stable and the balance cannot display it. Prevent vibration and drafts. Press the UNITS key to return to the weighing mode.

If you cannot cancel an error or other errors occurred, request service from the store where you purchased the balance or from your local A&D dealer.

35

Image 37
Contents 1WMPD4001917 Compliance with FCC rules This manual and MarksContents Maintenance Unpacking IntroductionPart Names and Functions NET indicatorSetting up your balance Setting UPPower source Breeze break Installing batteriesStorage Operation Turn the power on and OFFLCD backlight Conversion table UnitsSelecting a weighing unit NETBasic operation Selecting the counting mode Counting mode pcsCounting mode using the Acai function Storing the sample unit100 Percent mode %100 0 % 1000. %Selecting a comparison mode Setting exampleComparator Func001234 Cp Hi 000000Cp lo 000000 001230Calibration Calibration using a weightCal 060000Glp Cal 50000Gravity acceleration correction Cal 97985Functions Entering the function setting modeCp Hi Key operationACai Func PoffEnd Cp Hi End id UnitStoring weighing units Function list Umin Func ACaiSmpl LdinOptions EJ-03 InstallationEJ-02 USB Interface EJ-03 RS-232C serial interfaceEIA RS-232C Interface specificationsData format Data output modeCommand list Command modeEJ-07 / EJ-08 Underhook EJ-07 / EJ-08 InstallationEJ-11 Breeze break EJ-12 Carrying caseEJ-13 Density Determination Kit EJ-13 InstallationDensity specific gravity measurement Change the function table 000Example of density measurement 1751 8801 2000ID Number and GLP Setting the ID numberEnd id func Display character setData format for calibration report Output exampleData format for calibration test report Title Calibration test using a weightEnd Block Title block and End block20000 End 20000Error codes MaintenanceOverload error Range over notice Unit weight error Sample quantity noticeCal e -Cal e Error1Lb0 Hb0 Hb1 Lb1EJ series SpecificationsOther weighing units Options DimensionsValues of gravity at various locations Gravity Acceleration MAPWorld map