A&D EJ-3000, EJ-6100, EJ-440 Comparator, Func, Poff, Setting example, Selecting a comparison mode

Page 14

6. COMPARATOR

The results of the comparison are indicated by HI, OK or LO on the display. The comparison is as follows:

LO < Lower limit value OK Upper limit value < HI

Operating conditions (see the function setting “Cp”):

‰No comparison (comparator function disabled).

‰Compares all data.

‰Compares all stable data.

‰Compares plus data except those near zero (plus data greater than +4d).

‰Compares stable plus data except those near zero (stable plus data greater than +4d).

‰Compares all data except those near zero (all data greater than +4d or less than -4d).

‰Compares stable data except those near zero (stable data greater than +4d or less than -4d).

d = the smallest display division. e.g.: 4d = four display divisions

The upper limit and lower limit numerical values are common to each of the weighing, counting and percent mode. The example for EJ-120/200/300/410/610 is as follows.

Upper limit value “001010”: “10.10g” “1010pcs” “101.0%”

Lower limit value “000990”: “9.90g” “990pcs” “99.0%”

6-1. Setting example

Press and hold

This example will be “Compares plus data except

 

those near zero”.

func

Selecting a comparison mode

 

1.

Press and hold the

SAMPLE

key to display

 

 

func

.

 

 

 

 

 

 

 

 

 

 

 

(If the comparison mode is already set, press the

 

 

SAMPLE

 

key to go to “Entering the upper and

 

 

lower limit

values”.)

 

 

 

 

 

 

2.

Press the

 

 

 

key, then the balance displays

 

PRINT

 

 

poff X

 

.

 

 

 

 

 

 

 

 

 

3.

Press the

 

 

 

key several times to display

 

SAMPLE

 

 

Cp

X

.

 

 

 

 

 

 

 

 

 

4.

Press the

 

 

 

key several times to display

 

RE-ZERO

 

 

 

Cp

3

.

 

 

 

 

 

 

 

 

 

 

5.

Press the

 

 

 

key to store the settings.

 

PRINT

 

 

Cp Hi

appears

after

end

 

.

poff 0

Press several times

Cp 0

Press several times

Cp 3

To store

12

Image 14
Contents 1WMPD4001917 This manual and Marks Compliance with FCC rulesContents Maintenance Introduction UnpackingNET indicator Part Names and FunctionsPower source Setting UPSetting up your balance Storage Installing batteriesBreeze break LCD backlight Turn the power on and OFFOperation Units Conversion tableBasic operation NETSelecting a weighing unit Counting mode using the Acai function Counting mode pcsSelecting the counting mode Storing the sample unit100 0 % Percent mode %100 1000. %Comparator Setting exampleSelecting a comparison mode FuncCp lo 000000 Cp Hi 000000001234 001230Cal Calibration using a weightCalibration 060000Cal 50000 GlpCal 97985 Gravity acceleration correctionCp Hi Entering the function setting modeFunctions Key operationEnd Cp Hi Func PoffACai Storing weighing units UnitEnd id Function list Smpl Func ACaiUmin LdinEJ-02 USB Interface EJ-03 InstallationOptions EJ-03 RS-232C serial interfaceInterface specifications EIA RS-232CData output mode Data formatCommand mode Command listEJ-11 Breeze break EJ-07 / EJ-08 InstallationEJ-07 / EJ-08 Underhook EJ-12 Carrying caseDensity specific gravity measurement EJ-13 InstallationEJ-13 Density Determination Kit Example of density measurement 000Change the function table 2000 1751 8801End id func Setting the ID numberID Number and GLP Display character setData format for calibration test report Output exampleData format for calibration report End Block Calibration test using a weightTitle Title block and End block20000 20000 EndOverload error MaintenanceError codes Range over notice Unit weight error Sample quantity noticeLb0 Hb0 Error1Cal e -Cal e Hb1 Lb1Specifications EJ seriesOther weighing units Dimensions OptionsGravity Acceleration MAP Values of gravity at various locationsWorld map