A&D EJ-6100, EJ-440, EJ-3000, EJ-200, EJ-4100 Net, Selecting a weighing unit, Basic operation

Page 11

5-4. Selecting a weighing unit

Press the UNITS key to select a unit.

The following sections are a description of the three common units: g (gram mode), pcs (counting mode), and % (percent mode).

Each pressing switches the units available in the order described on the previous page.

5-5. Basic operation

1.Select a weighing unit.

2.When the display does not show zero, press the

RE-ZERO key to set the display to zero.

3.When using a tare (container), place the container on the weighing pan, and press the

RE-ZERO key to subtract the tare weight.

4.Place the object to be weighed on the pan or in the container.

Wait for the STABLE indicator to be displayed and read the value.

5.Remove the object from the pan.

‰The RE-ZERO key will zero the balance if the weight is within ±2% of the weighing capacity around the power-on zero point. The ZERO indicator turns on. When the weight exceeds +2% of the weighing capacity, it will be subtracted to zero as a tare weight. In this case the ZERO and NET marks turn on.

ZERO 00. STABLE indicator

Container

Weighing pan

NET

00.

ZERO

 

NET 1503.

STABLE indicator

Precautions during operation

‰Make sure that the STABLE indicator is on whenever reading or storing a value.

‰Do not press the keys with a sharp object such as a pencil.

‰Do not apply a shock or a load to the pan that is beyond the weighing capacity.

‰Keep the balance free from foreign objects such as dust or liquid.

‰Calibrate the balance periodically to keep weighing accuracy. (See “7. CALIBRATION”.)

9

Image 11
Contents 1WMPD4001917 Compliance with FCC rules This manual and MarksContents Maintenance Unpacking IntroductionPart Names and Functions NET indicatorPower source Setting UPSetting up your balance Storage Installing batteriesBreeze break LCD backlight Turn the power on and OFFOperation Conversion table UnitsBasic operation NETSelecting a weighing unit Storing the sample unit Counting mode pcsSelecting the counting mode Counting mode using the Acai function1000. % Percent mode %100 100 0 %Func Setting exampleSelecting a comparison mode Comparator001230 Cp Hi 000000001234 Cp lo 000000060000 Calibration using a weightCalibration CalGlp Cal 50000Gravity acceleration correction Cal 97985Key operation Entering the function setting modeFunctions Cp HiEnd Cp Hi Func PoffACai Storing weighing units UnitEnd id Function list Ldin Func ACaiUmin SmplEJ-03 RS-232C serial interface EJ-03 InstallationOptions EJ-02 USB InterfaceEIA RS-232C Interface specificationsData format Data output modeCommand list Command modeEJ-12 Carrying case EJ-07 / EJ-08 InstallationEJ-07 / EJ-08 Underhook EJ-11 Breeze breakDensity specific gravity measurement EJ-13 InstallationEJ-13 Density Determination Kit Example of density measurement 000Change the function table 1751 8801 2000Display character set Setting the ID numberID Number and GLP End id funcData format for calibration test report Output exampleData format for calibration report Title block and End block Calibration test using a weightTitle End Block20000 End 20000Range over notice Unit weight error Sample quantity notice MaintenanceError codes Overload errorHb1 Lb1 Error1Cal e -Cal e Lb0 Hb0EJ series SpecificationsOther weighing units Options DimensionsValues of gravity at various locations Gravity Acceleration MAPWorld map